Your browser doesn't support javascript.
loading
Cause of irreducible dislocation of a re-revision THR.
Murphy, Evelyn Patricia; Fenelon, Christopher; Russell, Shane; Condon, Finbarr.
Afiliação
  • Murphy EP; Department of Trauma and Orthopaedics, Limerick University Hospital, Limerick, Ireland.
  • Fenelon C; Limerick Univeristy Hospital, Limerick, Ireland.
  • Russell S; Department of Trauma and Orthopaedics, Limerick University Hospital, Limerick, Ireland.
  • Condon F; Department of Trauma and Orthopaedics, Limerick University Hospital, Limerick, Ireland.
BMJ Case Rep ; 20182018 Jun 04.
Article em En | MEDLINE | ID: mdl-29866666
Instability or recurrent dislocations are a common reason for revision of total hip replacements (THRs). Dual-mobility constructs can help improve stability by increasing the femoral head jump distance. These constructs are used to decrease the risk of re-revision in the setting of recurrent dislocation. This case describes an unusual case of irreducible dislocation of a re-revision THR due to intraprosthetic dislocation. The patient required open reduction and revision of this construct to a tripolar implant. This is important to appreciate from an emergency medicine point of view as repeated attempts at reduction will be fruitless and may result in a femur fracture. This particular type of dislocation is very rare. Perioperative considerations should include early referral to the orthopaedic team for reduction in theatre, and having a low threshold for open reduction. Revision options should be available when bringing a case like this to the operating theatre.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Falha de Prótese / Artroplastia de Quadril / Luxação do Quadril / Prótese de Quadril Limite: Humans / Male / Middle aged Idioma: En Revista: BMJ Case Rep Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Irlanda País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Falha de Prótese / Artroplastia de Quadril / Luxação do Quadril / Prótese de Quadril Limite: Humans / Male / Middle aged Idioma: En Revista: BMJ Case Rep Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Irlanda País de publicação: Reino Unido