Transcatheter valve-in-valve implantation in a degenerated very small Mitroflow prosthesis.
Lopez, Stéphane; Meyer, Pierre; Teboul, Jacques; Deharo, Pierre; Mihoubi, Alain; Elbeze, Jean Pierre; Lena, Diane; Camarasa, Philippe; Laborde, Jean Claude; Drogoul, Laurent.
Afiliação
  • Lopez S; Department of Cardiac Surgery, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Meyer P; Department of Cardiology, Cardiac Catheterization Laboratory, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Teboul J; Department of Cardiac Surgery, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Deharo P; Department of Cardiology, Cardiac Catheterization Laboratory, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Mihoubi A; Department of Cardiology, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Elbeze JP; Department of Cardiology, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Lena D; Department of Anesthesiology, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Camarasa P; Department of Anesthesiology, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Laborde JC; Department of Cardiology, Cardiac Catheterization Laboratory, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
  • Drogoul L; Department of Cardiology, Cardiac Catheterization Laboratory, Heart Institute Arnault Tzanck, CHU de Nice, Saint Laurent du Var, France.
Interact Cardiovasc Thorac Surg ; 27(6): 850-855, 2018 12 01.
Article em En | MEDLINE | ID: mdl-29917088
ABSTRACT

OBJECTIVES:

To assess the feasibility and results of 'valve-in-valve' implantation using the 23-mm CoreValve for the treatment of degenerated 19-mm and 21-mm Mitroflow bioprostheses.

METHODS:

We retrospectively analysed all consecutive patients who underwent transcatheter aortic valve implantation for 19-mm and 21-mm Mitroflow bioprostheses. The height of implantation with respect to the ring of the Mitroflow prosthesis was targeted at -6 mm for the first 3 cases. In the following cases, the target was higher to promote free supra-annular movement of the leaflets.

RESULTS:

The procedure was successful in 17 of 18 patients (94%). For implantations above the limit of -6 mm, the mean gradient was 10.4 ± 2.6 mmHg compared with 28.1 ± 11.6 mmHg for implantations below the limit of -6 mm (P < 0.01). For patients with severe stenosis as main mechanism of failure of the bioprosthesis, the mean post-procedural gradient was 31.2 ± 11.8 mmHg compared with 12.7 ± 6 mmHg in the absence of severe stenosis (P < 0.01). Patient-prosthesis mismatch (indexed effective orifice area ≤ 0.85 cm2/m2) and severe mismatch (indexed effective orifice area ≤ 0.65 cm2/m2) were present in 83% (15 of 18) and 27% (5 of 18) of patients, respectively. We did not notice any complications following the procedures. Six months after the procedure, functional status was improved in all patients.

CONCLUSIONS:

Our short series demonstrates the ability to perform transcatheter 'valve-in-valve' implantation in 19-mm and 21-mm Mitroflow prostheses with satisfactory results, but high post-procedural gradients and patient-prosthesis mismatch remain a relatively frequent problem mostly when severe stenosis is the main mechanism of failure. Implantation in a high position is critical to decrease the rate of high postimplantation gradients.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Bioprótese / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Interact Cardiovasc Thorac Surg Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Bioprótese / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Interact Cardiovasc Thorac Surg Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: França