Interchangeability of Procalcitonin Measurements Using the Point of Care Testing i-CHROMATM Reader and the Automated Liaison XL.
Clin Lab
; 64(6): 1097-1100, 2018 Jun 01.
Article
em En
| MEDLINE
| ID: mdl-29945312
BACKGROUND: Our aim was to verify if procalcitonin (PCT) measurements using the new point-of-care testing i-CHROMATM are interchangeable with those of Liaison XL. METHODS: One hundred seventeen serum samples were processed sequentially on a Liaison XL and i-CHROMATM. Statistical analysis was done using the Passing-Bablok regression, Bland-Altman test, and Cohen's Kappa statistic. RESULTS: Proportional and constant differences were observed between i-CHROMATM and Liaison XL. The 95% CI of the mean bias% was very large, exceeding the maximum allowable TE% and the clinical reference change value. However, the concordance between methods at the clinical relevant cutoffs was strong, with the exception of the 0.25 ng/mL cutoff which was moderate. CONCLUSIONS: Our data suggest that i-CHROMATM is not interchangeable with Liaison XL. However, while the strong concordance at the clinical relevant cutoffs allows us to consider i-CHROMATM a suitable option to Liaison XL to support clinicians' decision-making; nevertheless, the moderate agreement at the 0.25 ng/mL cutoff recommends caution in interpreting the data around this cutoff.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Automação Laboratorial
/
Testes Imediatos
/
Pró-Calcitonina
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
Limite:
Humans
Idioma:
En
Revista:
Clin Lab
Assunto da revista:
TECNICAS E PROCEDIMENTOS DE LABORATORIO
Ano de publicação:
2018
Tipo de documento:
Article
País de publicação:
Alemanha