Reliability of Hand-Held Transcranial Doppler with M-mode Ultrasound in Middle Cerebral Artery Measurement.
J Med Ultrasound
; 25(2): 76-81, 2017.
Article
em En
| MEDLINE
| ID: mdl-30065464
ABSTRACT
PURPOSE:
To determine the intra- and interrater agreement of mean flow velocity (MFV) and pulsatility index (PI) measurement in middle cerebral arteries, assessed by transcranial Doppler (TCD) with M-mode.METHODS:
Masked experienced neurosonologists performed TCD with M-mode using handheld probe in healthy adult volunteers. The Bland-Altman method for concordance and intraclass correlation coefficient were used.RESULTS:
Seventy-seven healthy volunteers and seven raters participated (3 on regular TCD shift and 4 off-shift). The intrarater absolute mean difference between measurements was 5.5 cm/s [95% confidence interval (CI), 4.7-6.3] for MVF and 0.073 (95% CI, 0.063-0.083) for PI. The difference between MFV measurements was significantly higher in off-shift raters (p = 0.015). The interrater absolute mean difference between measurements was 6.5 cm/s (95% CI, 5.5-7.5) for MVF and 0.065 (95% CI, 0.059-0.071) for PI. No influence was found for the middle cerebral artery side, volunteer's sex, or age, and there was no significant difference between raters. The intraclass correlation coefficient was 82.2% (95% CI 77.8-85.6) and 72.9% (95% CI 67.4-77.6) for MFV and PI, respectively.CONCLUSIONS:
There exists good intra- and interrater agreement in MFV and PI measurements using M-mode TCD. These results support the use of this noninvasive tool and are important for clinical and investigational purposes.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
J Med Ultrasound
Ano de publicação:
2017
Tipo de documento:
Article
País de afiliação:
Chile