MEASUREMENT OF PATIENT SKIN DOSE DISTRIBUTIONS IN THREE LEBANESE INTERVENTIONAL CARDIOLOGY SUITES.
Radiat Prot Dosimetry
; 183(3): 375-385, 2019 May 01.
Article
em En
| MEDLINE
| ID: mdl-30165531
Using a mesh of 30 thermoluminescent dosemeters, adults' patient skin doses were measured for 99 coronary angiography (CA) and 89 percutaneous coronary interventions (PCI) performed in three Lebanese hospitals. Average peak skin dose (Dskin,max) were 152 mGy (range: 16-1144) for CAs and 576 mGy (range: 7-3361) for PCIs. While only four patients had a Dskin,max value exceeding the 2 Gy threshold for skin injuries, several patients had skin dose values above 1 Gy at several distinct locations proving that Dskin,max alone is not sufficient for repetitive procedures; 2D dose maps are required instead. Dskin,max correlated well with total air kerma-area product (PKA,T) for PCI in Hospitals 1 and 2 (R = 0.91 and 0.76, respectively) enabling the setup of an alert level at PKA,T = 240 and 210 Gy cm2, respectively, corresponding to a Dskin,max of 2 Gy. This was not possible for Hospital 3 due to weak correlations between Dskin,max and PKA,T.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Doses de Radiação
/
Pele
/
Radiografia Intervencionista
/
Angiografia Coronária
Tipo de estudo:
Observational_studies
Limite:
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
País/Região como assunto:
Asia
Idioma:
En
Revista:
Radiat Prot Dosimetry
Ano de publicação:
2019
Tipo de documento:
Article
País de afiliação:
Líbano
País de publicação:
Reino Unido