Characterization of Stimulation Artifact Behavior in Simultaneous Electrocorticography Grid Stimulation and Recording.
Annu Int Conf IEEE Eng Med Biol Soc
; 2018: 4748-4751, 2018 Jul.
Article
em En
| MEDLINE
| ID: mdl-30441410
Bi-directional brain-computer interfaces (BCIs) require simultaneous stimulation and recording to achieve closed-loop operation. It is therefore important that the interface be able to distinguish between neural signals of interest and stimulation artifacts. Current bi-directional BCIs address this problem by temporally multiplexing stimulation and recording. This approach, however, is suboptimal in many BCI applications. Alternative artifact mitigation methods can be devised by investigating the mechanics of artifact propagation. To characterize stimulation artifact behaviors, we collected and analyzed electrocorticography (ECoG) data from eloquent cortex mapping. Ratcheting and phase-locking of stimulation artifacts were observed, as well as dipole-like properties. Artifacts as large as ±1,100 µV appeared as far as 15-37 mm away from the stimulating channel when stimulating at 10 mA. Analysis also showed that the majority of the artifact power was concentrated at the stimulation pulse train frequency (50 Hz) and its super-harmonics (100, 150, 200 Hz). Lower frequencies (0-32 Hz) experienced minimal artifact contamination. These findings could inform the design of future bi-directional ECoG-based BCIs.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Eletrocorticografia
Idioma:
En
Revista:
Annu Int Conf IEEE Eng Med Biol Soc
Ano de publicação:
2018
Tipo de documento:
Article
País de publicação:
Estados Unidos