Durotomy Surrogate and Seals for Intradural Spinal Cord Stimulators: Apparatus and Review of Clinical Methods and Materials.
Nagel, Sean J; Helland, Logan; Woodroffe, Royce W; Frizon, Leonardo A; Holland, Marshall T; Machado, Andre G; Yamaguchi, Satoshi; Gillies, George T; Howard, Matthew A; Wilson, Saul.
Afiliação
  • Nagel SJ; Neurological Surgery, Cleveland Clinic, Cleveland, OH, USA.
  • Helland L; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
  • Woodroffe RW; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
  • Frizon LA; Neurological Surgery, Cleveland Clinic, Cleveland, OH, USA.
  • Holland MT; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
  • Machado AG; Neurological Surgery, Cleveland Clinic, Cleveland, OH, USA.
  • Yamaguchi S; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
  • Gillies GT; Department of Mechanical and Aerospace Engineering, University of Virginia, Charlottesville, VA, USA.
  • Howard MA; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
  • Wilson S; Department of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City, IA, USA.
Neuromodulation ; 22(8): 916-929, 2019 Dec.
Article em En | MEDLINE | ID: mdl-30632655
INTRODUCTION: We are developing a novel intradural spinal cord stimulator for treatment of neuropathic pain and spasticity. A key feature is the means by which it seals the dura mater to prevent leakage of cerebrospinal fluid (CSF). We have built and employed a test rig that enables evaluation of candidate seal materials. METHODS: To guide the design of the test rig, we reviewed the literature on neurosurgical durotomies. The test rig has a mock durotomy slot with a dural substitute serving as the surrogate dura mater and water as the CSF. The primary experimental goal was to evaluate the effectiveness of candidate gasket materials as seals against CSF leakage in an implanted prototype device, at both normal and super-physiologic pressures. A secondary goal was to measure the transmembrane flows in a representative dural substitute material, to establish its baseline aqueous transport properties. RESULTS: The seals prevented leakage of water at the implantation site over periods of ≈ ten days, long enough for the scar tissue to form in the clinical situation. The seals also held at water pressure transients approaching 250 mm Hg. The residual volumetric flux of water through the dura substitute membrane (Durepair®) was δVT /A ≈ 0.24 mm3 /min/cm2 , consistent with expectations for transport through the porous membrane prior to closure and equalization of internal/external pressures. CONCLUSIONS: We have demonstrated the workability of obtaining robust seal against leakage at the implantation site of a novel intradural stimulator using a custom-designed test rig. Extension of the method to in vivo testing in a large animal model will be the next step.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dura-Máter / Estimulação da Medula Espinal / Vazamento de Líquido Cefalorraquidiano Limite: Animals / Humans Idioma: En Revista: Neuromodulation Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dura-Máter / Estimulação da Medula Espinal / Vazamento de Líquido Cefalorraquidiano Limite: Animals / Humans Idioma: En Revista: Neuromodulation Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos