Initial clinical trial of pins coated with fibroblast growth factor-2-apatite composite layer in external fixation of distal radius fractures.
Yanagisawa, Yohei; Ito, Atsuo; Hara, Yuki; Mutsuzaki, Hirotaka; Murai, Shinji; Fujii, Kengo; Sogo, Yu; Hirose, Motohiro; Oyane, Ayako; Kobayashi, Fumiko; Yamazaki, Masashi.
Afiliação
  • Yanagisawa Y; Department of Emergency and Critical Care Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
  • Ito A; Health Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Central 6, 1-1-1 Higashi, Tsukuba, Ibaraki, 305-8566, Japan.
  • Hara Y; Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
  • Mutsuzaki H; Department of Orthopaedic Surgery, Ibaraki Prefectural University of Health Sciences, 4669-2 Ami, Ami-machi, Inashiki-gun, Ibaraki, 300-0394, Japan.
  • Murai S; Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
  • Fujii K; Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
  • Sogo Y; Health Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Central 6, 1-1-1 Higashi, Tsukuba, Ibaraki, 305-8566, Japan.
  • Hirose M; Health Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Central 6, 1-1-1 Higashi, Tsukuba, Ibaraki, 305-8566, Japan.
  • Oyane A; Nanomaterials Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Central 5, 1-1-1 Higashi, Tsukuba, Ibaraki, 305-8565, Japan.
  • Kobayashi F; Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
  • Yamazaki M; Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan.
J Orthop ; 16(1): 69-73, 2019.
Article em En | MEDLINE | ID: mdl-30662242
ABSTRACT

BACKGROUND:

Pin tract infection and loosening are major complications and challenges in the treatment of fractures by external fixation. To address this issue, we developed titanium pins coated with a fibroblast growth factor 2 (FGF-2)-apatite composite layer. The purpose of this initial clinical trial is to clarify the safety and feasibility of using these pins for the external fixation of distal radius fractures.

METHODS:

Unstable, displaced fractures of the distal radius that were medically suitable for external fixation were treated using external fixation pins coated and uncoated with an FGF-2-apatite composite layer. The coated pin group (n = 5) comprised 5 women (average age, 70.4 ±â€¯5.9 years), whereas the uncoated pin group (n = 10) comprised 8 women and 2 men (average age, 64.4 ±â€¯11.7 years). The average duration of external fixation was 40.8 ±â€¯1.3 and 41.6 ±â€¯2.1 days for the coated and uncoated pin groups, respectively.

RESULTS:

All patients achieved fracture union. One patient in the uncoated group had severe pin tract infection on the day of pin extraction. No pin loosening or difficulty in pin removal was observed in either group. Bacterial growth was present in 5% and 25% of the pin sites in the coated and uncoated groups, respectively (p = 0.059). No adverse events such as tumor formation were observed for more than 2 years after surgery in the coated pin group.

CONCLUSIONS:

This study clarified the safety and feasibility of using pins coated with an FGF-2-apatite composite layer for the external fixation of distal radius fractures.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Orthop Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Orthop Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Japão