Comparison of Two Techniques to Measure Optic Nerve Sheath Diameter in Patients at Risk for Increased Intracranial Pressure.
Agrawal, Ashish; Cheng, Roger; Tang, Julin; Madhok, Debbie Y.
Afiliação
  • Agrawal A; Department of Anesthesia and Perioperative Care, University of California, San Francisco, CA.
  • Cheng R; Department of Neurology, Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ.
  • Tang J; Department of Anesthesia and Perioperative Care, University of California, San Francisco, CA.
  • Madhok DY; Department of Emergency Medicine and Neurology, University of California, San Francisco, CA.
Crit Care Med ; 47(6): e495-e501, 2019 06.
Article em En | MEDLINE | ID: mdl-30882482
ABSTRACT

OBJECTIVES:

Intracranial pressure over 20 mm Hg is associated with poor neurologic prognosis, but measuring intracranial pressure directly requires an invasive procedure. Dilation of the optic nerve sheath on axial ultrasound of the eye has been correlated with elevated intracranial pressure, but optimal cutoffs have been inconsistent possibly related to the measurement technique. A coronal technique has been studied on healthy volunteers but not on patients with high intracranial pressure. We compared two measurement techniques (axial and coronal) in patients with suspected high intracranial pressure due to trauma, bleeding, tumor, or infection.

DESIGN:

Prospective blinded observational study.

SETTING:

Two tertiary referral center ICUs. PATIENTS Twenty adults admitted to the ICU at risk for increased intracranial pressure expected to receive invasive intracranial monitoring.

INTERVENTIONS:

Ultrasound measurements of the optic nerve sheath in axial and coronal views either averaged between eyes or the highest in either eye. MEASUREMENTS AND MAIN

RESULTS:

Coronal measurements showed less variability between each eye than axial measurements (mean difference 0.5 mm vs 1 mm; p = 0.03) and were associated with high intracranial pressure at first measurement and over 24 hours (area under the receiver operating characteristic range 0.7-0.8). Mean and highest axial measurements showed improved association with first (area under the receiver operating characteristic 0.87-0.94) and highest intracranial pressure measurement (area under the receiver operating characteristic 0.89-0.96) within 24 hours. A cutoff of highest axial measurement in either eye greater than 6.2 mm or mean axial measurement between eyes of 5.6 mm had a sensitivity of 100% in predicting high intracranial pressure over the following 24 hours.

CONCLUSIONS:

The highest axial measurement of optic nerve sheath diameter in either eye is the most predictive of patients with high intracranial pressure in our population. This comparison of measurement techniques has not previously been described and should be further explored to set test cutoffs for ultrasound of the optic nerve sheath diameter.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nervo Óptico / Ultrassonografia / Hipertensão Intracraniana Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nervo Óptico / Ultrassonografia / Hipertensão Intracraniana Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Canadá