Establishing the interchangeability of arterial stiffness but not endothelial function parameters in healthy individuals.
Perrault, Raissa; Omelchenko, Alexander; Taylor, Carla G; Zahradka, Peter.
Afiliação
  • Perrault R; Canadian Centre for Agri-Food Research in Health and Medicine, St. Boniface Albrechtsen Research Centre, 351 Tache Avenue, Winnipeg, MB, R2H 2A6, Canada.
  • Omelchenko A; Department of Physiology and Pathophysiology, Faculty of Health Sciences, University of Manitoba, Winnipeg, Canada.
  • Taylor CG; Canadian Centre for Agri-Food Research in Health and Medicine, St. Boniface Albrechtsen Research Centre, 351 Tache Avenue, Winnipeg, MB, R2H 2A6, Canada.
  • Zahradka P; Canadian Centre for Agri-Food Research in Health and Medicine, St. Boniface Albrechtsen Research Centre, 351 Tache Avenue, Winnipeg, MB, R2H 2A6, Canada.
BMC Cardiovasc Disord ; 19(1): 190, 2019 08 06.
Article em En | MEDLINE | ID: mdl-31387535
ABSTRACT

BACKGROUND:

Development of instruments capable of detecting early stage vascular disease has increased interest in employing arterial stiffness (e.g. pulse wave velocity (PWV), augmentation index (AIx)) and endothelial dysfunction (e.g. reactive hyperemia index (RHI)) to diagnose atherosclerotic disease before occurrence of a cardiovascular event. However, amongst the equipment designed for this purpose, there is insufficient information regarding each of these parameters to establish appropriate cutoffs to distinguish between healthy and unhealthy blood vessels. To address these limitations, the study was designed to establish the upper arterial stiffness and endothelial function thresholds in a healthy population, by comparing the outputs from different instruments capable of measuring PWV, AIx and RHI.

METHODS:

A systematic comparison of PWV, AIx and RHI was conducted to determine the inter-relationships between these parameters of vascular functionality. Outputs were obtained non-invasively using three instruments, the VP-1000 (VP), SphygmoCor (SC), and EndoPAT (EP), in 40 apparently healthy males and females.

RESULTS:

Correlations were found between the brachial-ankle PWV and radial-ankle PWV (by VP and SC), and PWV (VP) with AIx (SC). The interchangeability of these outputs was demonstrated by the Bland Altman test, making it feasible to extrapolate cut-offs for radial-ankle PWV and AIx equivalent to brachial-ankle PWV that signify healthy vessels. In contrast, RHI showed no association with AIx, suggesting these endothelial and arterial parameters are functionally distinct.

CONCLUSIONS:

It was concluded that it is possible to compare the vascular function outputs of different instruments and identify healthy from unhealthy vessels, even though the approaches for quantifying the underlying physiological processes may differ. In this way, non-invasive determination of arterial function could be a new paradigm for detecting existing early stage asymptomatic atherosclerotic disease in individuals using techniques that are amenable to the clinical setting.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Endotélio Vascular / Índice Tornozelo-Braço / Rigidez Vascular / Análise de Onda de Pulso Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Cardiovasc Disord Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Endotélio Vascular / Índice Tornozelo-Braço / Rigidez Vascular / Análise de Onda de Pulso Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Cardiovasc Disord Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Canadá