Your browser doesn't support javascript.
loading
Correlation between Argus II array-retina distance and electrical thresholds of stimulation is improved by measuring the entire array.
Xu, Lucy T; Rachitskaya, Aleksandra V; DeBenedictis, Meghan J; Bena, James; Morrison, Shannon; Yuan, Alex.
Afiliação
  • Xu LT; Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • Rachitskaya AV; Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • DeBenedictis MJ; Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • Bena J; Department of Quantitative Health Sciences, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • Morrison S; Department of Quantitative Health Sciences, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • Yuan A; Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, OH, USA.
Eur J Ophthalmol ; 31(1): 194-203, 2021 Jan.
Article em En | MEDLINE | ID: mdl-31680546
ABSTRACT

PURPOSE:

To describe two methods of measuring Argus II array-retina distance and to correlate array-retina distance to electrode stimulation thresholds.

METHODS:

This was a case series of eight patients implanted with the Argus II. Spectral domain-optical coherence tomography array-retina distance was measured by two methods and correlated to corresponding electrode thresholds (1) array-retina distance at each array corner and the largest array-retina distance and (2) using manual optical coherence tomography segmentation, the average array-retina distance was determined for each group of four electrodes. Patients 1-5 and 6-8 were analyzed separately due to a different threshold programming software.

RESULTS:

The Spearman's rank coefficient between array-retina distance and thresholds was -0.006 (p = 0.98) for patients 1-5, and 0.16 (p = 0.59) for patients 6-8 with the first method. The Spearman's rank coefficient was 0.25 (p < 0.001) for patients 1-5 and 0.36 (p < 0.001) for patients 6-8 with the second method.

CONCLUSION:

There is a positive correlation between array-retina distance and threshold measurements when measuring the entire array but not when using a faster measurement method of four corners and largest array-retina distance.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Retina / Limiar Sensorial / Percepção Visual / Retinose Pigmentar / Próteses Visuais Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Ophthalmol Assunto da revista: OFTALMOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Retina / Limiar Sensorial / Percepção Visual / Retinose Pigmentar / Próteses Visuais Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Ophthalmol Assunto da revista: OFTALMOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos