Center of Mass Calculation in Combination with MS/MS Allows Robust Identification of Single Amino Acid Polymorphisms in Clinical Measurements of Insulin-Like Growth Factor-1.
Maus, Anthony; Kemp, Jennifer; Milosevic, Dragana; Renuse, Santosh; Pandey, Akhilesh; Singh, Ravinder J; Grebe, Stefan K G.
Afiliação
  • Maus A; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Kemp J; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Milosevic D; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Renuse S; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Pandey A; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Singh RJ; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
  • Grebe SKG; Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology , Mayo Clinic , Rochester , Minnesota 55905 , United States.
J Proteome Res ; 19(1): 186-193, 2020 01 03.
Article em En | MEDLINE | ID: mdl-31736316
ABSTRACT
Insulin-like growth factor-1 (IGF-1) measurement by high-resolution accurate mass-mass spectrometry (HRAM-MS) is replacing IGF-1 immunoassays and allows for identification of single amino acid variants; by contrast, both normal and deleterious sequence variants might be missed by immunoassays or non-HRAM-MS methods. We have developed an intact molecule HRAM-MS method to identify IGF-1 variants, distinguishing them by a center of mass (COM) calculation, followed by various tandem-MS activation techniques (HCD, ETD, ETciD, EThcD, UVPD). We found single amino acid variants in 841 of 146 620 patient samples (0.57%). Most were benign (A67T, A70T). We also observed a pathogenic variant (V44M), likely pathogenic variants (A38V, V17M), and a likely benign variant (A67V). For 207 samples from unique patients with residual serum, the MS variant results were confirmed by cell-free DNA sequencing. Our approach allows accurate quantitative reporting of functional IGF-1 in the presence of single amino acid variants. The COM approach potentially enables omission of tandem-MS for known, common variants, while the combination of COM and tandem-MS allows accurate identification in all cases we encountered. This approach should be applicable to qualitative and quantitative analyses of other peptides/proteins in clinical and research settings and might lend itself to the characterization of other protein variations.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fator de Crescimento Insulin-Like I / Espectrometria de Massas em Tandem Tipo de estudo: Diagnostic_studies / Prognostic_studies / Qualitative_research Limite: Humans Idioma: En Revista: J Proteome Res Assunto da revista: BIOQUIMICA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fator de Crescimento Insulin-Like I / Espectrometria de Massas em Tandem Tipo de estudo: Diagnostic_studies / Prognostic_studies / Qualitative_research Limite: Humans Idioma: En Revista: J Proteome Res Assunto da revista: BIOQUIMICA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos