Physical characteristics of infant endotracheal tubes.
Mullett, M D; Withers, M R.
Afiliação
  • Mullett MD; Department of Pediatrics, West Virginia University Medical Center, Morgantown 26506.
J Perinatol ; 8(2): 124-6, 1988.
Article em En | MEDLINE | ID: mdl-3193262
A change in the incidence of post-intubation complications in infants was the stimulus for the evaluation of the deformability of two brands of endotracheal tubes. The stiffness of each tube was evaluated at room and body conditions in the longitudinal axis of the tubes as well as a cross sectional plan. Shiley brand endotracheal tubes sizes 2.5 to 4.0 were less deformable than Portex brand of the same sizes.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Intubação Intratraqueal Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Humans / Infant / Newborn Idioma: En Revista: J Perinatol Assunto da revista: PERINATOLOGIA Ano de publicação: 1988 Tipo de documento: Article País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Intubação Intratraqueal Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Humans / Infant / Newborn Idioma: En Revista: J Perinatol Assunto da revista: PERINATOLOGIA Ano de publicação: 1988 Tipo de documento: Article País de publicação: Estados Unidos