Your browser doesn't support javascript.
loading
Validation of MRI quantitative susceptibility mapping of superparamagnetic iron oxide nanoparticles for hyperthermia applications in live subjects.
Deh, Kofi; Zaman, Marjan; Vedvyas, Yogindra; Liu, Zhe; Gillen, Kelly McCabe; O' Malley, Padraic; Bedretdinova, Dina; Nguyen, Thanh; Lee, Richard; Spincemaille, Pascal; Kim, Juyoung; Wang, Yi; Jin, Moonsoo M.
Afiliação
  • Deh K; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Zaman M; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Vedvyas Y; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Liu Z; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Gillen KM; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • O' Malley P; Department of Urology, University of Florida, Gainesville, FL, 32610, USA.
  • Bedretdinova D; Urology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Nguyen T; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Lee R; Urology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Spincemaille P; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Kim J; Department of Advanced Materials Engineering, Kangwon National University, Samcheok, 245-711, South Korea.
  • Wang Y; Department of Radiology, Weill Cornell Medicine, New York, NY, 10065, USA.
  • Jin MM; Department of Biomedical Engineering, Cornell University, Ithaca, NY, 14853, USA.
Sci Rep ; 10(1): 1171, 2020 01 24.
Article em En | MEDLINE | ID: mdl-31980695
ABSTRACT
The use of magnetic fluid hyperthermia (MFH) for cancer therapy has shown promise but lacks suitable methods for quantifying exogenous irons such as superparamagnetic iron oxide (SPIO) nanoparticles as a source of heat generation under an alternating magnetic field (AMF). Application of quantitative susceptibility mapping (QSM) technique to prediction of SPIO in preclinical models has been challenging due to a large variation of susceptibility values, chemical shift from tissue fat, and noisier data arising from the higher resolution required to visualize the anatomy of small animals. In this study, we developed a robust QSM for the SPIO ferumoxytol in live mice to examine its potential application in MFH for cancer therapy. We demonstrated that QSM was able to simultaneously detect high level ferumoxytol accumulation in the liver and low level localization near the periphery of tumors. Detection of ferumoxytol distribution in the body by QSM, however, required imaging prior to and post ferumoxytol injection to discriminate exogenous iron susceptibility from other endogenous sources. Intratumoral injection of ferumoxytol combined with AMF produced a ferumoxytol-dose dependent tumor killing. Histology of tumor sections corroborated QSM visualization of ferumoxytol distribution near the tumor periphery, and confirmed the spatial correlation of cell death with ferumoxytol distribution. Due to the dissipation of SPIOs from the injection site, quantitative mapping of SPIO distribution will aid in estimating a change in temperature in tissues, thereby maximizing MFH effects on tumors and minimizing side-effects by avoiding unwanted tissue heating.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Férricos / Óxido Ferroso-Férrico / Nanopartículas / Hipertermia Induzida Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: Sci Rep Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Férricos / Óxido Ferroso-Férrico / Nanopartículas / Hipertermia Induzida Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: Sci Rep Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos