Your browser doesn't support javascript.
loading
Implementation and validation of ASL perfusion measurements for population imaging.
Warnert, Esther A H; Steketee, Rebecca M E; Vernooij, Meike W; Ikram, M Arfan; Vogel, Mika; Hernandez Tamames, Juan-Antonio; Kotek, Gyula.
Afiliação
  • Warnert EAH; Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, the Netherlands.
  • Steketee RME; Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, the Netherlands.
  • Vernooij MW; Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, the Netherlands.
  • Ikram MA; Department of Epidemiology, Erasmus MC, Rotterdam, the Netherlands.
  • Vogel M; Department of Epidemiology, Erasmus MC, Rotterdam, the Netherlands.
  • Hernandez Tamames JA; GE Global Research, Munich, Germany.
  • Kotek G; Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, the Netherlands.
Magn Reson Med ; 84(4): 2048-2054, 2020 10.
Article em En | MEDLINE | ID: mdl-32239745
ABSTRACT

PURPOSE:

Pseudocontinuous arterial spin labeling (pCASL) allows for noninvasive measurement of regional cerebral blood flow (CBF), which has the potential to serve as biomarker for neurodegenerative and cardiovascular diseases. This work aimed to implement and validate pCASL on the dedicated MRI system within the population-based Rotterdam Study, which was installed in 2005 and for which software and hardware configurations have remained fixed.

METHODS:

Imaging was performed on two 1.5T MRI systems (General Electric); (I) the Rotterdam Study system, and (II) a hospital-based system with a product pCASL sequence. An in-house implementation of pCASL was created on scanner I. A flow phantom and three healthy volunteers (<27 years) were scanned on both systems for validation purposes. The data of the first 30 participants (86 ± 4 years) of the Rotterdam Study undergoing pCASL scans on scanner I only were analyzed with and without partial volume correction for gray matter.

RESULTS:

The validation study showed a difference in blood flow velocity, sensitivity, and spatial coefficient of variation of the perfusion-weighted signal between the two scanners, which was accounted for during post-processing. Gray matter CBF for the Rotterdam Study participants was 52.4 ± 8.2 ml/100 g/min, uncorrected for partial volume effects of gray matter. In this elderly cohort, partial volume correction for gray matter had a variable effect on measured CBF in a range of cortical and sub-cortical regions of interest.

CONCLUSION:

Regional CBF measurements are now included to investigate novel biomarkers in the Rotterdam Study. This work highlights that when it is not feasible to purchase a novel ASL sequence, an in-house implementation is valuable.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Circulação Cerebrovascular Limite: Aged / Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Circulação Cerebrovascular Limite: Aged / Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda