Comparative repeatability of pancreatic lipase assays in the commercial and in-house laboratory environments.
Cridge, Harry; Mackin, Andrew J; Lidbury, Jonathan A; Suchodolski, Jan S; Steiner, Jörg M.
Afiliação
  • Cridge H; Department of Clinical Sciences, College of Veterinary Medicine, Mississippi State University, Mississippi State, Mississippi, USA.
  • Mackin AJ; Department of Clinical Sciences, College of Veterinary Medicine, Mississippi State University, Mississippi State, Mississippi, USA.
  • Lidbury JA; Gastrointestinal Laboratory, Department of Small Animal Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Texas A&M University, College Station, Texas, USA.
  • Suchodolski JS; Gastrointestinal Laboratory, Department of Small Animal Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Texas A&M University, College Station, Texas, USA.
  • Steiner JM; Gastrointestinal Laboratory, Department of Small Animal Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Texas A&M University, College Station, Texas, USA.
J Vet Intern Med ; 34(3): 1150-1156, 2020 May.
Article em En | MEDLINE | ID: mdl-32249988
ABSTRACT

BACKGROUND:

Sensitivity and specificity for commercial and in-house pancreatic lipase immunoreactivity (cPLI) assays have been reported, but repeatability under routine clinical conditions is unknown.

OBJECTIVES:

To determine Coefficient of variation (CV) between replicates of a commercial assay (Spec cPL) and 2 in-house assays (VetScan cPL, Vcheck cPL) under routine conditions. Effects of sample condition or personnel on results. Potential directional bias between assays. ANIMALS Serum from 12 canine clinical patients.

METHODS:

Prospective study. Serum Spec cPL, VetScan cPL, and Vcheck cPL (6 aliquots each) were measured, and CVs were calculated, effects of sample condition and personnel were assessed using a linear mixed model, and direction of bias was assessed using least square mean cPLI concentration.

RESULTS:

Mean %CVs for Spec cPL, VetScan cPL, and Vcheck cPL were 5.5, 17.0, and 23.7%. Three of 6 VetScan cPL samples and 5/9 Vcheck cPL samples had an unacceptably high %CV (>20%). Transportation (Spec cPL) and sample condition or personnel (VetScan cPL, Vcheck cPL) did not affect repeatability. Least square mean cPL was higher for Spec cPL (807.9 µg/L) than for VetScan cPL (558.5 µg/L) or Vcheck cPL (399.8 µg/L). CONCLUSIONS AND CLINICAL IMPORTANCE For clinical use, the commercial Spec cPL has the highest repeatability, and Vcheck cPL has significantly lower repeatability. Both in-house assays evaluated may provide discrepant categorical results ("pancreatitis" versus "equivocal" versus "not pancreatitis") for the same sample. In-house pancreatic lipase concentrations may be lower than those determined by the Spec cPL assay.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pancreatite / Doenças do Cão / Lipase Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Animals Idioma: En Revista: J Vet Intern Med Assunto da revista: MEDICINA INTERNA / MEDICINA VETERINARIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pancreatite / Doenças do Cão / Lipase Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Animals Idioma: En Revista: J Vet Intern Med Assunto da revista: MEDICINA INTERNA / MEDICINA VETERINARIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos