Your browser doesn't support javascript.
loading
Implementation of an Anaesthesia Quality Improvement Programme to Reduce Fibreoptic Bronchoscope Repair Incidents.
Lu, Hsiao-Feng; Hung, Kuo-Chuan; Chiang, Min-Hsien; Yang, Johnson Chia-Shen; Luo, Sheng-Dean; Chin, Jo-Chi; Wang, Chih-Hsien; Sun, Cheuk-Kwan; Wu, Shao-Chun.
Afiliação
  • Lu HF; Department of Anaesthesiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Hung KC; Department of Anaesthesiology, Chi Mei Medical Center, Tainan, Taiwan.
  • Chiang MH; Department of Anaesthesiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Yang JC; Division of Plastic and Reconstructive Surgery, Department of Surgery, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Luo SD; Department of Otolaryngology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Chin JC; Department of Anaesthesiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Wang CH; Department of Anaesthesiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
  • Sun CK; Department of Emergency Medicine, E-Da Hospital, School of Medicine for International Students, I-Shou University, Kaohsiung, Taiwan.
  • Wu SC; Department of Anaesthesiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan.
Biomed Res Int ; 2020: 1091239, 2020.
Article em En | MEDLINE | ID: mdl-32337218
ABSTRACT

BACKGROUND:

This study was aimed at investigating the effectiveness of the implementation of a comprehensive quality improvement programme (QIP) for reducing the repair rate of the fibreoptic bronchoscope (FOB).

METHODS:

A three-stage improvement strategy was implemented between January 2013 and December 2016. Stage one is the acquisition of information on violations of practice guidelines, repair rate, cost of repair, and incidence of unavailability of FOB during anaesthesia induction of the previous year through auditing. Stage two is the implementation of a quality improvement campaign (QIC) based on the results of stage one. Stage three is the programme perpetuation through monitoring compliance with policy on FOB use by regular internal audits. The effectiveness was retrospectively analyzed on a yearly basis.

RESULTS:

The annual repair rate, repair cost, and incidence of FOB unavailability before the QIP implementation were 1%, 18,757 USD, and 1.4%, respectively. After QIC, the repair rate in 2013 dropped by 81% (from 1% in 2012 to 0.19% in 2013, p < 0.05). The annual repair cost fell by 32% from 18,758 USD (2012) to 12,820 USD (2013). Besides, the incidence of FOB unavailability plummeted by 71% from 1.4% to 0.4% during the same period. The annual repair rates and incidence of FOB unavailability remained lower in subsequent three years than those before QIP implementation.

CONCLUSION:

Implementation of a quality improvement programme was effective for reducing the rate and cost of FOB repair as well as unavailability rate, highlighting its beneficial impact on cost-effectiveness and patient safety in a tertiary referral center setting.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Broncoscópios / Falha de Equipamento / Melhoria de Qualidade / Manutenção Tipo de estudo: Guideline / Observational_studies / Risk_factors_studies / Sysrev_observational_studies Aspecto: Implementation_research Limite: Humans Idioma: En Revista: Biomed Res Int Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Taiwan

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Broncoscópios / Falha de Equipamento / Melhoria de Qualidade / Manutenção Tipo de estudo: Guideline / Observational_studies / Risk_factors_studies / Sysrev_observational_studies Aspecto: Implementation_research Limite: Humans Idioma: En Revista: Biomed Res Int Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Taiwan