Analytical performance evaluation of Sysmex CS-2500 and Stago STA Compact.
Yis, Ozgur M; Bugdayci, Guler; Pehlivan, Mine B; Yildiz, Rabia N; Alisik, Murat.
Afiliação
  • Yis OM; Department of Medical Biochemistry, Faculty of Medicine, Bolu Abant Izzet Baysal University.
  • Bugdayci G; Department of Medical Biochemistry, Faculty of Medicine, Bolu Abant Izzet Baysal University.
  • Pehlivan MB; Department of Medical Biochemistry, Faculty of Medicine, Bolu Abant Izzet Baysal University.
  • Yildiz RN; Department of Biochemistry, Bolu Izzet Baysal State Hospital, Bolu, Turkey.
  • Alisik M; Department of Medical Biochemistry, Faculty of Medicine, Bolu Abant Izzet Baysal University.
Blood Coagul Fibrinolysis ; 31(5): 324-329, 2020 Jul.
Article em En | MEDLINE | ID: mdl-32398465
ABSTRACT
The Sysmex Coagulation System-2500 (CS-2500) is a fully automated coagulation analyzer that uses the optical reaction method. In this study, we aimed to evaluate performance characteristics of the CS-2500 in two coagulation tests [prothrombin time (PT) and activated partial thromboplastin time (aPTT)] at our hospital laboratory. PT and aPTT measurements were performed using the CS-2500 and STA-Compact Diagnostica Stago coagulometers (STA-Compact). Then, precision, accuracy, reference range verification, and method comparison statistics were performed. In the precision study, which was performed with normal and pathologic controls for the PT-international normalized ratio (INR) and aPTT tests, all coefficient of variation% were found less than 3.5%. In the comparison study, the Passing-Bablok regression analysis demonstrated the good agreement between each analyzer for PT-INR (y = -0.081 + 1.07x and r = 0.962) and for aPTT (y = 5.498 + 0.86x and r = 0.944). Both analyzers exhibited less than 9.9% bias in the accuracy study. The reference range verification analyses revealed that the manufacturer ranges were acceptable. The verification studies of the CS-2500 coagulation system were acceptable; however, in the comparison studies, there were small differences between STA-Compact. Overall, we propose that these differences could be eliminated in future standardization studies performed to use the same reference ranges for all systems.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tempo de Tromboplastina Parcial / Tempo de Protrombina Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Blood Coagul Fibrinolysis Assunto da revista: ANGIOLOGIA / HEMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tempo de Tromboplastina Parcial / Tempo de Protrombina Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Blood Coagul Fibrinolysis Assunto da revista: ANGIOLOGIA / HEMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article
...