Your browser doesn't support javascript.
loading
Development of Volumetric Independent Dose Calculation System for Verification of the Treatment Plan in Image-Guided Adaptive Brachytherapy.
Kang, Sang-Won; Chung, Jin-Beom; Kim, Kyeong-Hyeon; Park, Ji-Yeon; Park, Hae-Jin; Cho, Woong; Olberg, Sven; Suh, Tae Suk; Park, Justin C.
Afiliação
  • Kang SW; Department of Biomedical Engineering, Department of Biomedicine and Health Sciences, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Chung JB; College of Medicine, Research Institute of Biomedical Engineering, The Catholic University of Korea, Seoul, South Korea.
  • Kim KH; Department of Radiation Oncology, Seoul National University Bundang Hospital, Seongnam, South Korea.
  • Park JY; Department of Biomedical Engineering, Department of Biomedicine and Health Sciences, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Park HJ; College of Medicine, Research Institute of Biomedical Engineering, The Catholic University of Korea, Seoul, South Korea.
  • Cho W; Department of Radiation Oncology, University of Florida Health Proton Therapy Institute, Jacksonville, FL, United States.
  • Olberg S; Department of Radiation Oncology, Ajou University School of Medicine, Suwon, South Korea.
  • Suh TS; Department of Radiation Oncology, Seoul National University Boramae Medical Center, Seoul, South Korea.
  • Park JC; Department of Biomedical Engineering, Washington University in St. Louis, St. Louis, MO, United States.
Front Oncol ; 10: 609, 2020.
Article em En | MEDLINE | ID: mdl-32477931
ABSTRACT

Purpose:

This study aimed to develop a volumetric independent dose calculation (vIDC) system for verification of the treatment plan in image-guided adaptive brachytherapy (IGABT) and to evaluate the feasibility of the vIDC in clinical practice with simulated cases.

Methods:

The vIDC is based on the formalism of TG-43. Four simulated cases of cervical cancer were selected to retrospectively evaluate the dose distributions in IGABT. Some reference point doses, such as points A and B and rectal points, were calculated by vIDC using absolute coordinate. The 3D dose volume was also calculated to acquire dose-volume histograms (DVHs) with grid resolutions of 1.0 × 1.0 (G1.0), 2.5 × 2.5 (G2.5), and 0.5 × 0.5 mm2 (G0.5). Dosimetric parameters such as D90% and D2cc doses covering 90% of the high-risk critical target volume (HR-CTV) and 2 cc of the organs at risk (OARs) were obtained from DVHs. D90% also converted to equivalent dose in 2-Gy fractions (EQD2) to produce the same radiobiological effect as external beam radiotherapy. In addition, D90% was obtained in two types with or without the applicator volume to confirm the effect of the applicator itself. Validation of the vIDC was also performed using gamma evaluation by comparison with Monte Carlo simulation.

Results:

The average percentage difference of point doses was <2.28%. The DVHs for the HR-CTV and OARs showed no significant differences between the vIDC and the treatment planning system (TPS). Without considering the applicator volume, the D90% of the HR-CTV calculated by the vIDC decreases with a decreasing calculated dose-grid size (32.4, 5.65, and -2.20 cGy in G2.5, G1.0, and G0.5, respectively). The overall D90% is higher when considering the applicator volume. The converted D90% by EQD2 ranged from -1.29 to 1.00%. The D2cc of the OARs showed that the averaged dose deviation is <10 cGy regardless of the dose-grid size. Based on gamma analysis, the passing rate was 98.81% for 3%/3-mm criteria.

Conclusion:

The vIDC was developed as an independent dose verification system for verification of the treatment plan in IGABT. We confirmed that the vIDC is suitable for second-check dose validation of the TPS under various conditions.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Front Oncol Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Coréia do Sul