Validation of T2 -based oxygen extraction fraction measurement with 15 O positron emission tomography.
Magn Reson Med
; 85(1): 290-297, 2021 01.
Article
em En
| MEDLINE
| ID: mdl-32643207
ABSTRACT
PURPOSE:
To evaluate the accuracy of T2 -based whole-brain oxygen extraction fraction (OEF) estimation by comparing it with gold standard 15 O-PET measurements.METHODS:
Sixteen healthy adult subjects underwent MRI and 15 O-PET OEF measurements on the same day. On MRI, whole-brain OEF was quantified by T2 -relaxation-under-spin-tagging (TRUST) MRI, based on subject-specific hematocrit. The TRUST OEF was compared to the whole-brain averaged OEF produced by 15 O-PET. Agreement between TRUST and 15 O-PET whole-brain OEF measurements was examined in terms of intraclass correlation coefficient (ICC) and in absolute OEF values. In a subset of 10 subjects, test-retest reproducibility of whole-brain OEF was also evaluated and compared between the two modalities.RESULTS:
Across the 16 subjects, the mean whole-brain OEF of TRUST and 15 O-PET were 36.44 ± 4.07% and 36.45 ± 3.65%, respectively, showing no difference between the two modalities (P = .99). TRUST whole-brain OEF strongly correlated with that of 15 O-PET (N = 16, ICC = 0.90, P = 4 × 10-7 ). The coefficient-of-variation of TRUST and 15 O-PET whole-brain OEF measurements were 1.79 ± 0.67% and 2.06 ± 1.55%, respectively, showing no difference between the two modalities (N = 10, P = .64). Further analyses on the effect of hematocrit revealed that correlation between PET OEF and TRUST OEF with assumed hematocrit remained significant (ICC = 0.8, P < 2 × 10-5 ).CONCLUSION:
Whole-brain OEF measured by TRUST was in excellent agreement with gold standard 15 O-PET, with highly comparable accuracy and reproducibility. These findings suggest that TRUST MRI can provide accurate quantification of whole-brain OEF noninvasively.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Circulação Cerebrovascular
/
Tomografia por Emissão de Pósitrons
Limite:
Adult
/
Humans
Idioma:
En
Revista:
Magn Reson Med
Assunto da revista:
DIAGNOSTICO POR IMAGEM
Ano de publicação:
2021
Tipo de documento:
Article
País de afiliação:
Estados Unidos