Your browser doesn't support javascript.
loading
Variation in intraocular lens calcification under different environmental conditions in eyes with supplementary sulcus-supported lenses.
Yildirim, Timur M; Khoramnia, Ramin; Schickhardt, Sonja K; Munro, Donald J; Merz, Patrick R; Son, Hyeck-Soo; Lieberwirth, Ingo; Auffarth, Gerd U.
Afiliação
  • Yildirim TM; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Khoramnia R; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Schickhardt SK; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Munro DJ; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Merz PR; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Son HS; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
  • Lieberwirth I; Deparment of Physical Chemistry of Polymers, Max Planck Institute for Polymer Research, Mainz, Germany.
  • Auffarth GU; The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital of Heidelberg, Heidelberg, Germany.
Am J Ophthalmol Case Rep ; 19: 100797, 2020 Sep.
Article em En | MEDLINE | ID: mdl-32760851
ABSTRACT

PURPOSE:

Analysis of explanted intraocular lenses (IOLs) from pseudophakic eyes with supplementary sulcus-supported IOLs.

METHODS:

In this laboratory investigation, ten supplementary and capsular bag IOLs were analyzed. All lenses were received between January 2012 and March 2018. Explants were examined morphologically with histological and electron microscopic techniques and patients' medical history was evaluated. Additionally, we used a technique new to this field Transmission Electron Microscopy and electron diffraction pattern analysis was performed to investigate the structure of the opacifying crystals in detail.

RESULTS:

Eleven lenses were explanted due to IOL opacification from seven polypseudophakic eyes In three cases the supplementary lens calcified, in three cases the capsular bag IOL (both lenses analyzed) and in one case both IOLs (only the supplementary was received). Additional surgical procedures and comorbidities included pars plana vitrectomy or Descemet stripping endothelial keratoplasty and diabetes mellitus. For each opacified lens, a varying layer of a Calcium phosphate beneath the optic surface was apparent. Crystal characterization revealed its composition to be Hydroxyapatite. CONCLUSIONS AND IMPORTANCE We report on a series of secondary calcification in lenses explanted from polypseudophakic eyes. In some cases, calcification occurred in the capsular bag lens, in other cases in the supplementary lens, or in both. The severity of the morphological change could be related to the comorbidities and the presence of surgery subsequent to the lens implantations. Detailed morphology of the opacifying crystals was revealed.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Am J Ophthalmol Case Rep Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha