Short single-wedge stems have higher risk of periprosthetic fracture than other cementless stem designs in Dorr type A femurs: a finite element analysis.
Nandi, Sumon; Shah, Anoli; Joukar, Amin; Becker, Kevin; Crutchfield, Cody; Goel, Vijay.
Afiliação
  • Nandi S; Department of Orthopaedics, University of Maryland School of Medicine, Baltimore, MD, USA.
  • Shah A; Department of Bioengineering, University of Toledo College of Engineering, Toledo, OH, USA.
  • Joukar A; Purdue University School of Mechanical Engineering, West Lafayette, IN, USA.
  • Becker K; Department of Bioengineering, University of Toledo College of Engineering, Toledo, OH, USA.
  • Crutchfield C; Department of Bioengineering, University of Toledo College of Engineering, Toledo, OH, USA.
  • Goel V; Department of Bioengineering, University of Toledo College of Engineering, Toledo, OH, USA.
Hip Int ; 32(3): 298-303, 2022 May.
Article em En | MEDLINE | ID: mdl-32865039
BACKGROUND: The use of total hip arthroplasty (THA) femoral stems that transmit force in a favourable manner to the femur may minimise periprosthetic fractures. Finite element analysis (FEA) is a computerised method that analyses the effect of forces applied to a structure with complex shape. Our aim was to apply FEA to compare primary THA cementless stem designs and their association with periprosthetic fracture risk. METHODS: 3-dimensional (3D) models of a Dorr Type A femur and 5 commonly used primary THA cementless stem designs (short single wedge, standard-length single wedge, modular, double-wedge metaphyseal filling, and cylindrical fully coated) were developed using computed tomography (CT) imaging. Implant insertion, single-leg stance, and twisting with a planted foot were simulated. FEA was performed, and maximum femoral strain along the implant-bone interface recorded. RESULTS: Femoral strain was highest with short single-wedge stem design (0.3850) and lowest with standard-length single-wedge design (0.0520). Location of maximum femoral strain varied by stem design, but not with implant insertion, single-leg stance, or twisting with a planted foot. Strain was as high during implant insertion as with single-leg stance or twisting with a planted foot. CONCLUSIONS: Our results suggest the risk of intraoperative and postoperative periprosthetic fracture with THA in a Dorr A femur is highest with short single-wedge stems and lowest with standard-length single-wedge stems. Consideration may be given to minimising the use of short single-wedge stems in THA. Implant-specific sites of highest strain should be carefully inspected for fracture.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Fraturas Periprotéticas / Fraturas do Fêmur / Prótese de Quadril Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Hip Int Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Fraturas Periprotéticas / Fraturas do Fêmur / Prótese de Quadril Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Hip Int Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos