Your browser doesn't support javascript.
loading
Validation of a Web-Based Planning Tool for Percutaneous Cryoablation of Renal Tumors.
van Oostenbrugge, Tim J; Heidkamp, Jan; Moche, Michael; Weir, Phil; Mariappan, Panchatcharam; Flanigan, Ronan; Pollari, Mika; Payne, Stephen; Kolesnik, Marina; Jenniskens, Sjoerd F M; Fütterer, Jurgen J.
Afiliação
  • van Oostenbrugge TJ; Department of Urology, Radboud University Medical Center, P.O. box 9101, 6500 HB, Nijmegen, The Netherlands. tim.vanoostenbrugge@radboudumc.nl.
  • Heidkamp J; Department of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Moche M; Department of Diagnostic and Interventional Radiology, University Hospital, Leipzig, Germany.
  • Weir P; NUMA Engineering Services Ltd, Dublin, Ireland.
  • Mariappan P; NUMA Engineering Services Ltd, Dublin, Ireland.
  • Flanigan R; Department of Mathematics, Indian Institute of Technology, Tirupati, India.
  • Pollari M; NUMA Engineering Services Ltd, Dublin, Ireland.
  • Payne S; Department of Neuroscience and Biomedical Engineering, Aalto University, Espoo, Finland.
  • Kolesnik M; Department of Engineering Science, Institute of Biomedical Engineering, University of Oxford, Oxford, UK.
  • Jenniskens SFM; Fraunhofer - FIT - Institute for Applied Information Technology, Sankt-Augustin, Germany.
  • Fütterer JJ; Department of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands.
Cardiovasc Intervent Radiol ; 43(11): 1661-1670, 2020 Nov.
Article em En | MEDLINE | ID: mdl-32935141
ABSTRACT

PURPOSE:

To validate a simulation environment for virtual planning of percutaneous cryoablation of renal tumors. MATERIALS AND

METHODS:

Prospectively collected data from 19 MR-guided procedures were used for validation of the simulation model. Volumetric overlap of the simulated ablation zone volume (Σ) and the segmented ablation zone volume (S; assessed on 1-month follow-up scan) was quantified. Validation metrics were DICE Similarity Coefficient (DSC; the ratio between twice the overlapping volume of both ablation zones divided by the sum of both ablation zone volumes), target overlap (the ratio between the overlapping volume of both ablation zones to the volume of S; low ratio means S is underestimated), and positive predictive value (the ratio between the overlapping volume of both ablation zones to the volume of Σ; low ratio means S is overestimated). Values were between 0 (no alignment) and 1 (perfect alignment), a value > 0.7 is considered good.

RESULTS:

Mean volumes of S and Σ were 14.8 cm3 (± 9.9) and 26.7 cm3 (± 15.0), respectively. Mean DSC value was 0.63 (± 0.2), and ≥ 0.7 in 9 cases (47%). Mean target overlap and positive predictive value were 0.88 (± 0.11) and 0.53 (± 0.24), respectively. In 17 cases (89%), target overlap was ≥ 0.7; positive predictive value was ≥ 0.7 in 4 cases (21%) and < 0.6 in 13 cases (68%). This indicates S is overestimated in the majority of cases.

CONCLUSION:

The validation results showed a tendency of the simulation model to overestimate the ablation effect. Model adjustments are necessary to make it suitable for clinical use.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Internet / Criocirurgia / Imagem por Ressonância Magnética Intervencionista / Neoplasias Renais Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Cardiovasc Intervent Radiol Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Internet / Criocirurgia / Imagem por Ressonância Magnética Intervencionista / Neoplasias Renais Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Cardiovasc Intervent Radiol Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda