Results of a project to calibrate mercury sphygmomanometer blood pressure-measuring devices in Egypt.
Ghareeb, Soliman; Youssef, Ghada; Ghareeb, Haytham Soliman; El-Mageed, Hazem Abd; Mesalm, Muhammad H; Talaat, Remon; Eltawil, Alaaeldin; Hasan, Doaa M; Ibrahim, M Mohsen.
Afiliação
  • Ghareeb S; Cardiology Department, Cairo University, Cairo, Egypt.
  • Youssef G; Cardiology Department, Cairo University, Cairo, Egypt. ghadayoussef@kasralainy.edu.eg.
  • Ghareeb HS; Cardiology Department, El Fayoum University, El Fayoum, Egypt.
  • El-Mageed HA; Cardiology Department, Cairo University, Cairo, Egypt.
  • Mesalm MH; Cardiology Department, Cairo University, Cairo, Egypt.
  • Talaat R; Ipsos Healthcare, Cairo, Egypt.
  • Eltawil A; National Institute of Standards (NIS), Cairo, Egypt.
  • Hasan DM; Cardiology Department, Cairo University, Cairo, Egypt.
  • Ibrahim MM; Cardiology Department, Cairo University, Cairo, Egypt.
J Hum Hypertens ; 35(10): 921-926, 2021 10.
Article em En | MEDLINE | ID: mdl-33037321
Hypertension (HTN) is a common health problem and a major cardiovascular risk factor. Accurate measurement of blood pressure (BP) is mandatory for proper diagnosis and follow-up. The aim of this study was to evaluate the accuracy of mercury sphygmomanometer BP-measuring devices in public hospitals in Cairo, Egypt. Fifty public hospitals were included, and 10% of all mercury sphygmomanometer devices in each hospital were tested. Assessment included physical condition (e.g., mercury status, lid of the device, state of the rubber tubes), leakage rate, and calibration accuracy (as compared with a reference device). Devices were approved as accurate when they could successfully pass all three assessment tests. The total number of sphygmomanometer devices was 465. The overall pass rate was 1.3% (six devices). Twenty-five (5.2%) devices passed all of the physical tests, 50 (10.8%) passed the leakage test, and 50 (16.5%) passed the calibration accuracy test. There were 162 (34.8%) devices that showed a high leakage rate (>80 mmHg) and thus were not tested for calibration accuracy. In conclusion, most of the mercury sphygmomanometer devices in hospitals are neglected and not checked regularly for any errors. A plan should be made to gradually replace those failed devices with new, validated, and well-calibrated devices, preferably devices that do not contain mercury.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Mercúrio Tipo de estudo: Risk_factors_studies Limite: Humans País/Região como assunto: Africa Idioma: En Revista: J Hum Hypertens Assunto da revista: ANGIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Egito País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Mercúrio Tipo de estudo: Risk_factors_studies Limite: Humans País/Região como assunto: Africa Idioma: En Revista: J Hum Hypertens Assunto da revista: ANGIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Egito País de publicação: Reino Unido