Your browser doesn't support javascript.
loading
Transcatheter Aortic Valve Thrombogenesis: A Foreign Materials Perspective.
Salim, Md Tausif; Sadri, Vahid; Nair, Priya; Schwaner, Daniel; Apte, Mugdha; Yoganathan, Ajit P.
Afiliação
  • Salim MT; School of Chemical and Biomolecular Engineering, Georgia Institute of Technology, 387 Technology Circle NW, Technology Enterprise Park, Suite 200, Atlanta, GA, 30313, USA.
  • Sadri V; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Nair P; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Schwaner D; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Apte M; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Yoganathan AP; School of Chemical and Biomolecular Engineering, Georgia Institute of Technology, 387 Technology Circle NW, Technology Enterprise Park, Suite 200, Atlanta, GA, 30313, USA. ajit.yoganathan@bme.gatech.edu.
Cardiovasc Eng Technol ; 12(1): 28-36, 2021 02.
Article em En | MEDLINE | ID: mdl-33277684
ABSTRACT

PURPOSE:

The initiation of thrombus formation in transcatheter aortic valves (TAVs) is not well understood. The foreign material components of a TAV may play a key role in TAV thrombogenesis. The goal of this study was to evaluate the thrombogenic potential of a TAV (entire valve) and its stent (with skirt).

METHODS:

Blood was collected from eight human donors with citrate anticoagulation and later reconstituted with calcium chloride. A low-volume steady flow loop (flow rate = 0.8 L/min) was designed to facilitate three separate conditions (experimental duration = 1 h) per donor blood (1) control (n = 8), (2) stent-with-skirt (leaflets removed from a 23 mm SAPIEN XT valve; n = 8) and (3) entire valve (an intact 23 mm SAPIEN XT valve; n = 8). Samples were collected at the start and end of each experiment. Serum D-Dimer and thrombin-antithrombin (TAT) concentrations were measured as markers of thrombogenicity.

RESULTS:

There was no significant change in serum D-Dimer and TAT concentration with time for the control group. An increasing trend in D-Dimer and TAT concentration was observed with time for the stent-with-skirt group. Interestingly, there was a decreasing trend in serum D-Dimer and TAT concentration with time for the entire valve (leaflet dominating) group. Moreover, changes in D-Dimer and TAT concentration were significantly different between the stent-with-skirt and entire valve (leaflet dominating) groups.

CONCLUSION:

Stent-with-skirt was found to impart the most prominent thrombogenic effect, indicating the significance of blood-stent and blood-skirt interactions in TAV thrombosis.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Limite: Humans Idioma: En Revista: Cardiovasc Eng Technol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Limite: Humans Idioma: En Revista: Cardiovasc Eng Technol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Estados Unidos