Your browser doesn't support javascript.
loading
Single-fraction prostate stereotactic body radiotherapy: Dose reconstruction with electromagnetic intrafraction motion tracking.
Jaccard, Maud; Ehrbar, Stefanie; Miralbell, Raymond; Hagen, Tobias; Koutsouvelis, Nikolaos; Poulsen, Per; Rouzaud, Michel; Tanadini-Lang, Stephanie; Tsoutsou, Pelagia; Guckenberger, Matthias; Zilli, Thomas.
Afiliação
  • Jaccard M; Radiation Oncology, University Hospital of Geneva, Switzerland. Electronic address: MJaccard@genolier.net.
  • Ehrbar S; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Switzerland.
  • Miralbell R; Radiation Oncology, University Hospital of Geneva, Switzerland; Radiation Oncology, Teknon Oncologic Institute, Barcelona, Spain; Faculty of Medicine, Geneva University, Geneva, Switzerland.
  • Hagen T; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Switzerland.
  • Koutsouvelis N; Radiation Oncology, University Hospital of Geneva, Switzerland.
  • Poulsen P; Department of Oncology and Danish Center for Particle Therapy, Aarhus University Hospital, Denmark.
  • Rouzaud M; Radiation Oncology, University Hospital of Geneva, Switzerland.
  • Tanadini-Lang S; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Switzerland.
  • Tsoutsou P; Radiation Oncology, University Hospital of Geneva, Switzerland; Faculty of Medicine, Geneva University, Geneva, Switzerland.
  • Guckenberger M; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Switzerland.
  • Zilli T; Radiation Oncology, University Hospital of Geneva, Switzerland; Faculty of Medicine, Geneva University, Geneva, Switzerland. Electronic address: Thomas.Zilli@hcuge.ch.
Radiother Oncol ; 156: 145-152, 2021 03.
Article em En | MEDLINE | ID: mdl-33310011
ABSTRACT

PURPOSE:

To reconstruct the dose delivered during single-fraction urethra-sparing prostate stereotactic body radiotherapy (SBRT) accounting for intrafraction motion monitored by intraprostatic electromagnetic transponders (EMT).

METHODS:

We analyzed data of 15 patients included in the phase I/II "ONE SHOT" trial and treated with a single fraction of 19 Gy to the planning target volume (PTV) and 17 Gy to the urethra planning risk volume. During delivery, prostate motion was tracked with implanted EMT. SBRT was interrupted when a 3-mm threshold was trespassed and corrected unless the offset was transient. Motion-encoded reconstructed (MER) plans were obtained by splitting the original plans into multiple sub-beams with isocenter shifts based on recorded EMT positions, mimicking prostate motion during treatment. We analyzed intrafraction motion and compared MER to planned doses.

RESULTS:

The median EMT motion range (±SD) during delivery was 0.26 ± 0.09, 0.22 ± 0.14 and 0.18 ± 0.10 cm in the antero-posterior, supero-inferior, and left-right axes, respectively. Treatment interruptions were needed for 8 patients because of target motion beyond limits in the antero-posterior (n = 6) and/or supero-inferior directions (n = 4). Comparing MER vs. original plan there was a median relative dose difference of -1.9% (range, -7.9 to -1.0%) and of +0.5% (-0.3-1.7%) for PTV D98% and D2%, respectively. The clinical target volume remained sufficiently covered with a median D98% difference of -0.3% (-1.6-0.5%). Bladder and rectum dosimetric parameters showed significant differences between original and MER plans, but mostly remained within acceptable limits.

CONCLUSIONS:

The dosimetric impact of intrafraction prostate motion was minimal for target coverage for single-fraction prostate SBRT with real-time electromagnetic tracking combined with beam gating.