Your browser doesn't support javascript.
loading
Applicability of Cork as Novel Modifiers to Develop Electrochemical Sensor for Caffeine Determination.
Monteiro, Mayra K S; Da Silva, Djalma R; Quiroz, Marco A; Vilar, Vítor J P; Martínez-Huitle, Carlos A; Dos Santos, Elisama V.
Afiliação
  • Monteiro MKS; Laboratório de Eletroquímica Ambiental e Aplicada, Universidade Federal do Rio Grande do Norte, Lagoa Nova 59.072-900, Brazil.
  • Da Silva DR; Laboratório de Eletroquímica Ambiental e Aplicada, Universidade Federal do Rio Grande do Norte, Lagoa Nova 59.072-900, Brazil.
  • Quiroz MA; Laboratório de Eletroquímica Ambiental e Aplicada, Universidade Federal do Rio Grande do Norte, Lagoa Nova 59.072-900, Brazil.
  • Vilar VJP; Grupo de Investigación en Energía y Ambiente, Universidad de las Américas Puebla, ExHda. Sta. Catarina Martir, Cholula 72820, Mexico.
  • Martínez-Huitle CA; Laboratory of Separation and Reaction Engineering-Laboratory of Catalysis and Materials (LSRE-LCM), Chemical Engineering Department, Faculty of Engineering, University of Porto, 4200-465 Porto, Portugal.
  • Dos Santos EV; Laboratório de Eletroquímica Ambiental e Aplicada, Universidade Federal do Rio Grande do Norte, Lagoa Nova 59.072-900, Brazil.
Materials (Basel) ; 14(1)2020 Dec 24.
Article em En | MEDLINE | ID: mdl-33374209
ABSTRACT
This study aims to investigate the applicability of a hybrid electrochemical sensor composed of cork and graphite (Gr) for detecting caffeine in aqueous solutions. Raw cork (RAC) and regranulated cork (RGC, obtained by thermal treatment of RAC with steam at 380 °C) were tested as modifiers. The results clearly showed that the cork-graphite sensors, GrRAC and GrRGC, exhibited a linear response over a wide range of caffeine concentration (5-1000 µM), with R2 of 0.99 and 0.98, respectively. The limits of detection (LOD), estimated at 2.9 and 6.1 µM for GrRAC and GrRGC, suggest greater sensitivity and reproducibility than the unmodified conventional graphite sensor. The low-cost cork-graphite sensors were successfully applied in the determination of caffeine in soft drinks and pharmaceutical formulations, presenting well-defined current signals when analyzing real samples. When comparing electrochemical determinations and high performance liquid chromatography measurements, no significant differences were observed (mean accuracy 3.0%), highlighting the potential use of these sensors to determine caffeine in different samples.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Materials (Basel) Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Materials (Basel) Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil