Validation of Bone Density and Microarchitecture Measurements of the Load-Bearing Femur in the Human Knee Obtained Using In Vivo HR-pQCT Protocol.
Keen, Christopher E; Whittier, Danielle E; Firminger, Colin R; Edwards, W Brent; Boyd, Steven K.
Afiliação
  • Keen CE; McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
  • Whittier DE; McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
  • Firminger CR; McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, Alberta, Canada; Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Alberta, Canada.
  • Edwards WB; McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, Alberta, Canada; Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Alberta, Canada.
  • Boyd SK; McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada; Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Alberta, Canada. El
J Clin Densitom ; 24(4): 651-657, 2021.
Article em En | MEDLINE | ID: mdl-33531205
High resolution peripheral quantitative computed tomography (HR-pQCT) was designed to study bone mineral density (BMD) and microarchitecture in peripheral sites at the distal radius and tibia. With the introduction of the second generation HR-pQCT scanner (XtremeCT II, Scanco Medical) that has a larger, longer gantry it is now possible to study the human knee in vivo using HR-pQCT. Previous validation of HR-pQCT measurements at the distal radius and tibia against micro-CT is not representative of the knee because the increased cross-sectional area, greater amount of soft tissue surrounding the scan region, and different imaging protocol result in potentially increased beam hardening effects and photon scatter and different signal-to-noise ratio. The objective of this study is to determine the accuracy of density and microarchitecture measurements in the human knee measured by HR-pQCT using an in vivo protocol. Twelve fresh-frozen cadaver knees were imaged using in vivo HR-pQCT (60.7 µm) protocol. Subsequentially, distal femurs were extracted and imaged using a higher resolution (30.3 µm) ex vivo protocol, replicating micro-CT imaging. Scans were registered so that agreement of density and bone microarchitecture measurements could be determined using linear regression and Bland-Altman plots. All density and microarchitecture outcomes were highly correlated between the 2 protocols (R2 > 0.89) albeit with statistically significant differences between absolute measures based on paired t tests. All parameters showed accuracy between 4.5% and 8.7%, and errors were highly systematic, particularly for trabecular BMD and trabecular thickness (R2 > 0.93). We found that BMD and microarchitecture measurements in the distal femur obtained using an in vivo HR-pQCT knee protocol contained systematic errors, and accurately represented measurements obtained using a micro-CT equivalent imaging protocol. This work establishes the validity and limitations of using HR-pQCT to study the BMD and microarchitecture of human knees in future clinical studies.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Densidade Óssea / Fêmur Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: J Clin Densitom Assunto da revista: ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Canadá País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Densidade Óssea / Fêmur Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: J Clin Densitom Assunto da revista: ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Canadá País de publicação: Estados Unidos