Clinical validation of time reduction strategy in continuous step-and-shoot mode during SPECT acquisition.
Picone, Valentin; Makris, Nikolaos; Boutevin, Fanny; Roy, Sarah; Playe, Margot; Soussan, Michael.
Afiliação
  • Picone V; GE Healthcare, 78530, 283 Rue de la Miniere, Buc, France.
  • Makris N; GE Healthcare, 78530, 283 Rue de la Miniere, Buc, France.
  • Boutevin F; GE Healthcare, 78530, 283 Rue de la Miniere, Buc, France.
  • Roy S; GE Healthcare, 78530, 283 Rue de la Miniere, Buc, France.
  • Playe M; Department of Nuclear Medicine, Avicenne Hospital, HUPSSD, APHP, Paris, France.
  • Soussan M; Department of Nuclear Medicine, Avicenne Hospital, HUPSSD, APHP, Paris, France. Michael.soussan@aphp.fr.
EJNMMI Phys ; 8(1): 10, 2021 Feb 02.
Article em En | MEDLINE | ID: mdl-33532876
ABSTRACT

BACKGROUND:

The SwiftScan solution (General Electric Healthcare) combines a new low-energy high-resolution sensitivity collimator and a tomographic step-and-shoot continuous (SSC) mode acquisition. The purpose of this study is to determine whether SSC mode can be used in clinical practice with shorter examination times, while preserving image quality and ensuring accurate semi-quantification. Twenty bone scan and 10 lung scan studies were randomly selected over a period of 2 months. Three sets of image datasets were produced step-and-shoot (SS) acquisition, simulated 25% count reduction using the Poisson resampling method (SimSS), and SimSS continuous acquisition (SimSSC), where SimSS was summed with counts acquired during detector head rotation. Visual assessment (5-point Likert scale, 2 readers) and semi-quantitative evaluation (50 focal uptake from 10 bone studies), assessed by SUVmean, coefficient of variation (COV), and contrast-to-noise ratio (CNR), were performed using t test and Bland-Altman analysis.

RESULTS:

Intra-reader agreement was substantial for reader 1 (k = 0.71) and for reader 2 (k = 0.61). Inter-reader agreement was substantial for SS set (k = 0.93) and moderate for SimSSC (k = 0.52). Bland-Altman analysis showed a good interchangeability of SS and SimSSC SUV values. The mean CNR between SS and SimSSC was not significantly different 42.9 ± 43.7 [23.7-62.1] vs. 43.1 ± 46 [22.9-63.3] (p = 0.46), respectively. COV values, assessing noise level, did not deviate significantly between SS and SimSSC 0.20 ± 0.08 [0.18-0.23] vs. 0.21 ± 0.08, [0.18-0.23] (p = 0.15), respectively, whereas a significant difference was demonstrated between SS and SimSS 0.20 ± 0.08 [0.18-0.23] vs. 0.23 ± 0.09 [0.20-0.25] (p < 0.0001), respectively.

CONCLUSIONS:

SSC mode acquisition decreases examination time by approximately 25% in bone and lung SPECT/CT studies compared to SS mode (~ 2 min per single-bed SPECT), without compromising image quality and signal quantification. This SPECT sensitivity improvement also offers the prospect of more comfortable exams, with less motion artifacts, especially in painful or dyspneic patients.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: EJNMMI Phys Ano de publicação: 2021 Tipo de documento: Article País de afiliação: França