Validation of a secondary dose check tool against Monte Carlo and analytical clinical dose calculation algorithms in VMAT.
Piffer, Stefano; Casati, Marta; Marrazzo, Livia; Arilli, Chiara; Calusi, Silvia; Desideri, Isacco; Fusi, Franco; Pallotta, Stefania; Talamonti, Cinzia.
Afiliação
  • Piffer S; Department of Experimental and Clinical Biomedical Sciences, University of Florence, Florence, Italy.
  • Casati M; National Institute of Nuclear Physics (INFN), Florence, Italy.
  • Marrazzo L; Department of Medical Physics, Careggi University Hospital, Florence, Italy.
  • Arilli C; Department of Medical Physics, Careggi University Hospital, Florence, Italy.
  • Calusi S; Department of Medical Physics, Careggi University Hospital, Florence, Italy.
  • Desideri I; Department of Experimental and Clinical Biomedical Sciences, University of Florence, Florence, Italy.
  • Fusi F; Department of Experimental and Clinical Biomedical Sciences, University of Florence, Florence, Italy.
  • Pallotta S; Department of Experimental and Clinical Biomedical Sciences, University of Florence, Florence, Italy.
  • Talamonti C; Department of Experimental and Clinical Biomedical Sciences, University of Florence, Florence, Italy.
J Appl Clin Med Phys ; 22(4): 52-62, 2021 Apr.
Article em En | MEDLINE | ID: mdl-33735491
PURPOSE: Patient-specific quality assurance (QA) is very important in radiotherapy, especially for patients with highly conformed treatment plans like VMAT plans. Traditional QA protocols for these plans are time-consuming reducing considerably the time available for patient treatments. In this work, a new MC-based secondary dose check software (SciMoCa) is evaluated and benchmarked against well-established TPS (Monaco and Pinnacle3 ) by means of treatment plans and dose measurements. METHODS: Fifty VMAT plans have been computed using same calculation parameters with SciMoCa and the two primary TPSs. Plans were validated with measurements performed with a 3D diode detector (ArcCHECK) by translating patient plans to phantom geometry. Calculation accuracy was assessed by measuring point dose differences and gamma passing rates (GPR) from a 3D gamma analysis with 3%-2 mm criteria. Comparison between SciMoCa and primary TPS calculations was made using the same estimators and using both patient and phantom geometry plans. RESULTS: TPS and SciMoCa calculations were found to be in very good agreement with validation measurements with average point dose differences of 0.7 ± 1.7% and -0.2 ± 1.6% for SciMoCa and two TPSs, respectively. Comparison between SciMoCa calculations and the two primary TPS plans did not show any statistically significant difference with average point dose differences compatible with zero within error for both patient and phantom geometry plans and GPR (98.0 ± 3.0% and 99.0 ± 3.0% respectively) well in excess of the typical 95 % clinical tolerance threshold. CONCLUSION: This work presents results obtained with a significantly larger sample than other similar analyses and, to the authors' knowledge, compares SciMoCa with a MC-based TPS for the first time. Results show that a MC-based secondary patient-specific QA is a clinically viable, reliable, and promising technique, that potentially allows significant time saving that can be used for patient treatment and a per-plan basis QA that effectively complements traditional commissioning and calibration protocols.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia de Intensidade Modulada Tipo de estudo: Guideline / Health_economic_evaluation Limite: Humans País/Região como assunto: Europa Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia de Intensidade Modulada Tipo de estudo: Guideline / Health_economic_evaluation Limite: Humans País/Região como assunto: Europa Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália País de publicação: Estados Unidos