Your browser doesn't support javascript.
loading
Simultaneous T2 and T2 mapping of multiple sclerosis lesions with radial RARE-EPI.
Herrmann, Carl J J; Els, Antje; Boehmert, Laura; Periquito, Joao; Eigentler, Thomas Wilhelm; Millward, Jason M; Waiczies, Sonia; Kuchling, Joseph; Paul, Friedemann; Niendorf, Thoralf.
Afiliação
  • Herrmann CJJ; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Els A; Department of Physics, Humboldt University of Berlin, Berlin, Germany.
  • Boehmert L; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Periquito J; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Eigentler TW; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Millward JM; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Waiczies S; Chair of Medical Engineering, Technical University of Berlin, Berlin, Germany.
  • Kuchling J; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Paul F; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany.
  • Niendorf T; Experimental and Clinical Research Center, a joint cooperation between the Charité Medical Faculty and the Max Delbrück Center for Molecular Medicine, Berlin, Germany.
Magn Reson Med ; 86(3): 1383-1402, 2021 09.
Article em En | MEDLINE | ID: mdl-33951214
ABSTRACT

PURPOSE:

The characteristic MRI features of multiple sclerosis (MS) lesions make it conceptually appealing to pursue parametric mapping techniques that support simultaneous generation of quantitative maps of 2 or more MR contrast mechanisms. We present a modular rapid acquisition with relaxation enhancement (RARE)-EPI hybrid that facilitates simultaneous T2 and T2∗ mapping (2in1-RARE-EPI).

METHODS:

In 2in1-RARE-EPI the first echoes in the echo train are acquired with a RARE module, later echoes are acquired with an EPI module. To define the fraction of echoes covered by the RARE and EPI module, an error analysis of T2 and T2∗ was conducted with Monte Carlo simulations. Radial k-space (under)sampling was implemented for acceleration (R = 2). The feasibility of 2in1-RARE-EPI for simultaneous T2 and T2∗ mapping was examined in a phantom study mimicking T2 and T2∗ relaxation times of the brain. For validation, 2in1-RARE-EPI was benchmarked versus multi spin-echo (MSE) and multi gradient-echo (MGRE) techniques. The clinical applicability of 2in1-RARE-EPI was demonstrated in healthy subjects and MS patients.

RESULTS:

There was a good agreement between T2 / T2∗ values derived from 2in1-RARE-EPI and T2 / T2∗ reference values obtained from MSE and MGRE in both phantoms and healthy subjects. In patients, MS lesions in T2 and T2∗ maps deduced from 2in1-RARE-EPI could be just as clearly delineated as in reference maps calculated from MSE/MGRE.

CONCLUSION:

This work demonstrates the feasibility of radially (under)sampled 2in1-RARE-EPI for simultaneous T2 and T2∗ mapping in MS patients.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esclerose Múltipla Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esclerose Múltipla Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha
...