Evaluation of the inter and intraobserver reproducibility of the GRASP method: a goniometric method to measure the isolated glenohumeral range of motion in the shoulder joint.
Ruiz Ibán, Miguel Angel; Alonso Güemes, Susana; Ruiz Díaz, Raquel; Asenjo Gismero, Cristina Victoria; Lorente Gomez, Alejandro; Diaz Heredia, Jorge.
Afiliação
  • Ruiz Ibán MA; Unidad de Hombro y Codo, Hospital Universitario Ramón y Cajal, Cta Colmenar km 9,100, 28046, Madrid, Spain. drmri@hotmail.com.
  • Alonso Güemes S; Unidad de Hombro y Codo, Hospital Universitario Ramón y Cajal, Cta Colmenar km 9,100, 28046, Madrid, Spain.
  • Ruiz Díaz R; Unidad de Hombro y Codo, Hospital Universitario Ramón y Cajal, Cta Colmenar km 9,100, 28046, Madrid, Spain.
  • Asenjo Gismero CV; Unidad de Hombro y Codo, Hospital FREMAP Majadahonda, Madrid, Spain.
  • Lorente Gomez A; Unidad de Hombro y Codo, Hospital Universitario Ramón y Cajal, Cta Colmenar km 9,100, 28046, Madrid, Spain.
  • Diaz Heredia J; Unidad de Hombro y Codo, Hospital Universitario Ramón y Cajal, Cta Colmenar km 9,100, 28046, Madrid, Spain.
J Exp Orthop ; 8(1): 37, 2021 May 15.
Article em En | MEDLINE | ID: mdl-33993351
ABSTRACT

PURPOSE:

To evaluate the intra and interobserver reproducibility of a new goniometric method for evaluating the isolated passive range of motion of the glenohumeral joint in an outpatient setting.

METHODS:

This is a prospective observational study on healthy subjects. The Glenohumeral ROM Assessment with Scapular Pinch (GRASP) method is a new method for assessing the isolated range of motion (ROM) of the glenohumeral joint (GH) by a single examiner with a clinical goniometer. It measures the isolated glenohumeral passive abduction (GH-AB), passive external rotation (GH-ER) and internal rotation (GH-IR) with the arm at 45º of abduction. These three GH ROM parameters were measured in both shoulders of 30 healthy volunteers (15 males/15 females, mean age41.6[SD = 10.3] years). The full shoulder passive abduction, passive external rotation and internal rotation 45º of abduction were measured by the same examiners with a goniometer for comparison. One examiner made two evaluations and a second examiner made a third one. The primary outcome was the intra- and interobserver reproducibility of the measurements assessed with intraclass correlation coefficients (ICC) and the Bland-Altman plot.

RESULTS:

The intra-observer ICC for isolated glenohumeral ROM were 0.84 ± 0.07 for GH-ABD, 0.63 ± 0.09 for GH-ER, and 0.61 ± 0.14 for GH-IR. The inter-observer ICC for isolated glenohumeral ROM were 0.86 ± 0.06 for GH-ABD, 0.68 ± 0.12 for GH-ER, and 0.62 ± 0.14 for GH-IR. These results were similar to those obtained for full shoulder ROM assessment with a goniometer.

CONCLUSION:

The GRASP method is reproducible for quick assessment of isolated glenohumeral ROM. LEVEL OF EVIDENCE III.