Single-shot pseudo-centric EPI for magnetization-prepared imaging.
Lee, Hyun-Soo; Hwang, Seon-Ha; Park, Jaeseok; Park, Sung-Hong.
Afiliação
  • Lee HS; MRI Laboratory, Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Korea.
  • Hwang SH; Siemens Healthineers Ltd, Seoul, Korea.
  • Park J; MRI Laboratory, Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Korea.
  • Park SH; Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Korea.
Magn Reson Med ; 86(5): 2656-2665, 2021 11.
Article em En | MEDLINE | ID: mdl-34184310
ABSTRACT

PURPOSE:

To implement a single-shot centric-reordered EPI (1sh-CenEPI), which reduces TE significantly, thus enabling to improve SNR for magnetization-prepared imaging.

METHODS:

We proposed a 1sh-CenEPI in which grouped oscillating readout gradients, phase-encoding blips within each group, and big phase-encoding jumps between two consecutive groups are incorporated to encode whole k-space from the center to the edges in a single shot. The concept was tested on phantoms and human brains at 3 T. In addition, the proposed reordering scheme was applied to pseudo-continuous arterial spin labeling for evaluating the efficiency of the centric reordering in magnetization-prepared imaging.

RESULTS:

The proposed 1sh-CenEPI reduced TE from 50 ms to 1.4 ms for gradient-echo EPI, and from 100 ms to 7 ms for spin-echo EPI, while the elongation of readout duration was below 10% of the whole readout duration in most cases. The 1sh-CenEPI images exhibited no distinct geometric distortion both in phantom and human brain, comparable to the conventional two-shot center-out EPI. In pseudo-continuous arterial spin labeling results, 3-fold temporal SNR increase and 2-fold spatial SNR increase in the perfusion-weighted images were achieved with 1sh-CenEPI compared with the conventional linear ordering, whereas the cerebral blood flow values were consistent with previous studies.

CONCLUSION:

The proposed 1sh-CenEPI significantly reduced TE while maintaining similar readout window and providing images comparable to the conventional linear and multishot center-out EPI images. It can be a qualified candidate as a new readout for various magnetization-prepared imaging techniques.