Your browser doesn't support javascript.
loading
Intraoperative detection of human meningioma using a handheld visible resonance Raman analyzer.
Zhang, Liang; Zhou, Yan; Wu, Binlin; Zhang, Shengjia; Zhu, Ke; Liu, Cheng-Hui; Yu, Xinguang; Alfano, Robert R.
Afiliação
  • Zhang L; Medical School of Nankai University, Tianjin, 300071, China.
  • Zhou Y; Department of Neurosurgery, General Hospital, Beijing, 100853, China.
  • Wu B; Department of Neurosurgery, Air Force Medical Center, Beijing, 100142, China.
  • Zhang S; Physics Department and CSCU Center for Nanotechnology, Southern Connecticut State University, New Haven, CT, 06515, USA.
  • Zhu K; JRME Co., Ltd, Taizhou, 225300, Jiangsu, China.
  • Liu CH; Institute of Physics, Chinese Academy of Sciences (CAS), PO Box 603, Beijing, 100190, China.
  • Yu X; Institute for Ultrafast Spectroscopy and Lasers, Department of Physics, The City College of the City University of New York, NY, 10031, New York, USA.
  • Alfano RR; Medical School of Nankai University, Tianjin, 300071, China. xinguang_yu@263.net.
Lasers Med Sci ; 37(2): 1311-1319, 2022 Mar.
Article em En | MEDLINE | ID: mdl-34365551
ABSTRACT
To report for the first time the preliminary results for the evaluation of a VRR-LRR™ analyzer based on visible resonance Raman technique to identify human meningioma grades and margins intraoperatively. Unprocessed primary and recurrent solid human meningeal tissues were collected from 33 patients and underwent Raman analysis during surgeries. A total of 1180 VRR spectra were acquired from fresh solid tissues using a VRR-LRR™ analyzer. A confocal HR Evolution (HORIBA, France SAS) Raman system with 532-nm excitation wavelength was also used to collect data for part of the ex vivo samples after they were thawed from - 80 °C for comparison. The preliminary analysis led to the following observations. (1) The intensity ratio of VRR peaks of protein to fatty acid (I2934/I2888) decreased with the increase of meningioma grade. (2) The ratio of VRR peaks of phosphorylated protein to amid I (I1588/I1639) decreased for the higher grade of meningioma. (3) Three RR vibration modes at 1378, 3174, and 3224 cm-1 which were related to the molecular vibrational bands of oxy-hemeprotein, amide B, and amide A protein significantly changed in peak intensities in the two types of meningioma tissues compared to normal tissue. (4) The changes in the intensities of VRR modes of carotenoids at 1156 and 1524 cm-1 were also found in the meningioma boundary. The VRR-LRR™ analyzer demonstrates a new approach for label-free, rapid, and objective identification of primary human meningioma in quasi-clinical settings. The accuracy for detecting meningioma tissues using support vector machines (SVMs) was over 70% based on Raman peaks of key biomolecules and up to 100% using principal component analysis (PCA).
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Meníngeas / Meningioma Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Lasers Med Sci Assunto da revista: BIOTECNOLOGIA / RADIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Meníngeas / Meningioma Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Lasers Med Sci Assunto da revista: BIOTECNOLOGIA / RADIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China