Your browser doesn't support javascript.
loading
New Linked-Wire-Type External Fixator (the Ichi-Fixator System) for Metacarpal Fractures: Case Series and Literature Review.
Igeta, Yuka; Ichihara, Satoshi; Hara, Akira; Suzuki, Masao; Otani, Satoshi; Maruyama, Yuichiro; Nozawa, Masahiko.
Afiliação
  • Igeta Y; Department of Orthopedics Surgery, Juntendo University Nerima Hospital, Tokyo, Japan.
  • Ichihara S; Hand Surgery Center, Juntendo University Urayasu Hospital, Chiba, Japan.
  • Hara A; Hand Surgery Center, Juntendo University Urayasu Hospital, Chiba, Japan.
  • Suzuki M; Hand Surgery Center, Juntendo University Urayasu Hospital, Chiba, Japan.
  • Otani S; Hand Surgery Center, Juntendo University Urayasu Hospital, Chiba, Japan.
  • Maruyama Y; Department of Orthopedic Surgery, Juntendo Urayasu Hospital, Chiba, Japan.
  • Nozawa M; Department of Orthopedics Surgery, Juntendo University Nerima Hospital, Tokyo, Japan.
J Hand Surg Asian Pac Vol ; 26(3): 403-409, 2021 Sep.
Article em En | MEDLINE | ID: mdl-34380408
ABSTRACT

Background:

External fixator is an effective treatment option in the comminuted or unstable metacarpal fractures. The new linked-wire type of external fixator (the Ichi-Fixator System) developed for hand and wrist fractures enables fine adjustment of the fixation under fluoroscopic inspection through small screws inside the fixator fitted externally. This technique is designed to reinforce the stability and rigidity of conventional percutaneous Kirchner wire fixation. The aim of this study is to assess the effectiveness of the fixator for comminuted or unstable metacarpal fractures through the evaluation of short-term results in ten cases.

Methods:

Patients were fixated with fixator pins and metal clamps using the Ichi-Fixator System. All patients were then examined for post-operative complications, functional recovery, visual analogue scale (VAS) score, and the Quick Disabilities of the Arm, Shoulder, and Hand (Q-DASH) score at the end of the final follow-up consultation.

Results:

Operative treatment using the Ichi-Fixator System facilitates anatomical reduction and allows immediate full mobilization, producing good outcomes. No infections were observed in all cases, including those at pin insertion sites. There were no implant failure and nonunion. Patients could perform all routine activities with normal grip strength and retained a full range of hand motion without pain. This treatment method with improved postoperative comfort would allow immediate return to work, which would clearly boost patient satisfaction.

Conclusions:

Ichi-Fixator System is considered to be useful for the treatment of metacarpal fractures, and this will provide a new surgical option for all hand fractures.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ossos Metacarpais Limite: Humans Idioma: En Revista: J Hand Surg Asian Pac Vol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ossos Metacarpais Limite: Humans Idioma: En Revista: J Hand Surg Asian Pac Vol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão