Evaluation of analytic and clinical performance of two immunoassays for detecting thyroid-stimulating receptor antibody in the diagnosis of Graves' disease.
Hu, Yao; Ni, Jiajin; Cen, Yi; Zhang, Buyue; Wu, Wenqing; Cheng, Wei; Huang, Mingying; Guan, Ming.
Afiliação
  • Hu Y; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Ni J; Department of Nursing, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Cen Y; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Zhang B; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Wu W; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Cheng W; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Huang M; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
  • Guan M; Department of Laboratory Medicine, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.
J Clin Lab Anal ; 35(12): e23950, 2021 Dec.
Article em En | MEDLINE | ID: mdl-34752648
ABSTRACT

OBJECTIVE:

To evaluate the analytical and clinical performance of two immunoassays for diagnosis of Graves' disease (GD), the Immulite thyroid-stimulating immunoglobulin (TSI), and Elecsys Anti-TSH receptor (TSHR) assay.

METHODS:

Precision and analytical measurement range were assessed using pooled samples of patients. The comparison between the two methods was evaluated using 579 clinical samples, and receiver operating characteristic (ROC) curves were drawn using the final diagnosis as reference. Clinical sensitivity and specificity, accuracy, positive predictive value (PPV), and negative predictive value (NPV) were calculated for the two tests.

RESULTS:

The repeatability and intermediate imprecision coefficient of variation (CV%) of the TSI assay were 3.8% and 4.1% at 0.95 IU/L, and 3.5% and3.6% at 19.5 IU/L, respectively. The assays were linear over a range 0.27-38.5 IU/L. There was a high correlation between the quantitative results of the two methods (correlation coefficient r = 0.930). The cut-off value obtained by ROC analysis for TSI assay was 0.7 IU/L with sensitivity of 93.7% and specificity of 85.1%. An overall qualitative agreement of 91.5% between two methods was observed. Among 44 patients with discordant qualitative results, the TSI assay provided more satisfactory results consistent with clinical diagnoses.

CONCLUSION:

The TSI assay showed excellent analytical performance and provided a high PPV for GD.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Receptores da Tireotropina / Imunoensaio / Doença de Graves / Imunoglobulinas Estimuladoras da Glândula Tireoide Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies / Qualitative_research Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Receptores da Tireotropina / Imunoensaio / Doença de Graves / Imunoglobulinas Estimuladoras da Glândula Tireoide Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies / Qualitative_research Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China