Your browser doesn't support javascript.
loading
Electromagnetic Exposure of Personnel Involved in Cardiac MRI Examinations in 1.5T, 3T and 7T Scanners.
Sklinda, Katarzyna; Karpowicz, Jolanta; Stepniewski, Andrzej.
Afiliação
  • Sklinda K; Department of Radiology, Centre of Postgraduate Medical Education, Marymoncka 99/103, 01-813 Warszawa, Poland.
  • Karpowicz J; Department of Bioelectromagnetics, Central Institute for Labour Protection-National Research Institute (CIOP-PIB), Czerniakowska 16, 00-701 Warszawa, Poland.
  • Stepniewski A; ECOTECH-COMPLEX Centre, University of Maria Curie-Sklodowska, Gleboka 39, 20-612 Lublin, Poland.
Article em En | MEDLINE | ID: mdl-35010336
ABSTRACT
(1)

Background:

It has been hypothesised that a significant increase in the use of cardiac magnetic resonance (CMR), for example, when examining COVID-19 convalescents using magnetic resonance imaging (MRI), has an influence the exposure profiles of medical personnel to static magnetic fields (STmf). (2)

Methods:

Static exposure to STmf (SEmf) was recorded during activities that modelled performing CMR by radiographers. The motion-induced time variability of that exposure (TVEmf) was calculated from SEmf samples. The results were compared with (i) labour law requirements; (ii) the distribution of vertigo perception probability near MRI magnets; and (iii) the exposure profile when actually performing a head MRI. (3)

Results:

The exposure profiles of personnel managing 42 CMR scans (modelled using medium (1.5T), high (3T) and ultrahigh (7T) field scanners) were significantly different than when managing a head MRI. The majority of SEmf and TVEmf samples (up to the 95th percentile) were at low vertigo perception probability (SEmf < 500 mT, TVEmf < 600 mT/s), but a small fraction were at medium/high levels; (4)

Conclusion:

Even under the "normal working conditions" defined for SEmf (STmf < 2T) by labour legislation (Directive 2013/35/EC), increased CMR usage increases vertigo-related hazards experienced by MRI personnel (a re-evaluation of electromagnetic safety hazards is suggested in the case of these or similar changes in work organisation).
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Exposição Ocupacional / COVID-19 Limite: Humans Idioma: En Revista: Int J Environ Res Public Health Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Polônia