Dual ambient plasma source ionization mass spectrometry for the rapid detection of trace sterols in urban water: Rapid detection of trace sterols in urban water.
J Mass Spectrom
; 57(2): e4809, 2022 Jan 06.
Article
em En
| MEDLINE
| ID: mdl-35075730
ABSTRACT
A direct analytical method based on dual ambient plasma ion source tandem mass spectrometry was used for the simultaneous determination of four sterols in the environment. This technology has very high sensitivity and the method detects the four sterols in methanol-water (13) solutions with limits of detection (LOD) and limits of quantification (LOQ) ranging from 1.2 ng/L to 6.9 ng/L and 7.6 ng/L to 10.0 ng/L, respectively. The method was also used to test water quality at three locations within the city and successfully detected all four sterols at very low concentrations. The dual plasma source tandem mass spectrometry technique is extremely simple, rapid, sensitive and highly efficient compared to other traditional methods, providing a useful screening tool for sterols in water.