Micra AV leadless pacemaker implantation after transcatheter aortic valve implantation.
Mechulan, Alexis; Prevot, Sébastien; Peret, Angélique; Nait-Saidi, Lyassine; Miliani, Ichem; Leong-Feng, Lauriane; Leude-Vaillant, Elisabeth; Vaillant, Alain; Cornen, Alain; Latiere, Bernard; Giacomoni, Marie-Paule; Collet, Frédéric; Bechet, Vincent; Bouharaoua, Ahmed; Dieuzaide, Pierre.
Afiliação
  • Mechulan A; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Prevot S; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Peret A; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Nait-Saidi L; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Miliani I; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Leong-Feng L; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Leude-Vaillant E; Ramsay Santé, Hôpital Privé Clairval, Service de Chirurgie Cardiaque, Marseille, France.
  • Vaillant A; Ramsay Santé, Hôpital Privé Clairval, Service de Chirurgie Cardiaque, Marseille, France.
  • Cornen A; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Latiere B; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Giacomoni MP; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Collet F; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Bechet V; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Bouharaoua A; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
  • Dieuzaide P; Ramsay Santé, Hôpital Privé Clairval, Service Cardiologie, Marseille, France.
Pacing Clin Electrophysiol ; 45(11): 1310-1315, 2022 11.
Article em En | MEDLINE | ID: mdl-35661380
ABSTRACT

BACKGROUND:

Transvenous pacemaker (PM) implantation is a complication in patients undergoing transcatheter aortic valve implantation (TAVI). Recently, a second generation of leadless PMs able of atrioventricular (AV) synchronous pacing has been introduced and could be an alternative when ventricular pacing is required after TAVI. Real-world data on Micra AV after TAVI are still lacking. Our aim was to determine the per- and post-procedural outcomes in patients with Micra AV leadless PM implantation after TAVI.

METHODS:

A total of 20 consecutive patients underwent Micra AV leadless PM implantation after TAVI between November 2020 and June 2021.

RESULTS:

The main indication for ventricular pacing was high-degree AV block (55% of patients) and left bundle branch block (LBBB) associated with prolonged HV interval (45% of patients). At discharge, mean (SD) ventricular pacing threshold was 0.397 ± 0.11 V at 0.24 ms and ventricular impedance was 709.4 ± 139.1 Ω. At 1-month follow-up, 95% of patients were programmed in VDD pacing mode. Mean (SD) ventricular pacing threshold was 0.448 ± 0.094 V at 0.24 ms. In patients with ventricular> pacing > 90% (n = 5), mean AM-VP was 72.5% ± 8.3%. Pacing threshold at 1 month was not significantly different compared to discharge (p = .1088). Mean (SD) impedance was 631.0 ± 111.9 Ω, which remained stable at discharge (p = .0813). No procedural complications occurred during implantation. At 1-month follow-up, two patients displayed atrial under-sensing.

CONCLUSIONS:

Micra AV leadless PM implantation after TAVI is associated with a low complication rate and good device performance at 1-month post-implantation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Bloqueio Atrioventricular / Substituição da Valva Aórtica Transcateter Limite: Humans Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Bloqueio Atrioventricular / Substituição da Valva Aórtica Transcateter Limite: Humans Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: França