One-year results of the GORE EXCLUDER Conformable AAA Endoprosthesis system in the United States regulatory trial.
Rhee, Robert; Oderich, Gustavo; Han, Sukgu; Long, Chandler; Muck, Patrick; Moore, Erin; Matsumura, Jon.
Afiliação
  • Rhee R; Maimonides Medical Center, Brooklyn, NY. Electronic address: rrhee@maimonidesmed.org.
  • Oderich G; UT Health, The University of Texas Health Science Center, Houston, TX.
  • Han S; Keck Hospital of University of Southern California, Los Angeles, CA.
  • Long C; Duke University Medical Center, Durham, NC.
  • Muck P; Good Samaritan Hospital, Cincinnati, OH.
  • Moore E; Baptist Medical Center, Jacksonville, FL.
  • Matsumura J; University of Wisconsin School of Medicine and Public Health, Madison, WI.
J Vasc Surg ; 76(4): 951-959.e2, 2022 10.
Article em En | MEDLINE | ID: mdl-35688351
ABSTRACT

OBJECTIVE:

To report the 1-year clinical outcomes from the GORE EXCLUDER Conformable AAA Endoprosthesis system in the US regulatory trial.

METHODS:

The study is a prospective, multicenter, investigational device exemption clinical trial at 31 US sites with core laboratory assessment of imaging and independent event adjudication. The primary safety (incidence of major adverse events at 30 days) and effectiveness end points (successful aneurysm treatment at 1 year) were assessed in a cohort of patients with abdominal aortic aneurysms (AAAs).

RESULTS:

We enrolled 80 patients between December 19, 2017, and February, 27, 2019. The mean maximum aortic diameter was 57.7 ± 7.95 mm (range, 42.5-82.7 mm) with an average patient age of 73.5 ± 8.14 years (range, 56-96 years). Overall technical success was 100% (80/80). The mean hospital length stay was 1.2 ± 0.6 days (range, 1-4 days). No primary safety end point events were observed, including no death, stroke, myocardial infarction, bowel ischemia, paraplegia, respiratory failure, renal failure, procedural blood loss of more than 1000 mL, or thromboembolic events including limb occlusion or distal emboli. There were no type I or III endoleaks detected on the 1-, 6-, or 12-month follow-up computed tomography scans. There were no stent fractures, device migrations (≥10 mm), AAA ruptures, or conversions to open surgical repair observed. Two patients had AAA sac growth of more than 5 mm at 1 year owing to type II endoleaks. There were no aneurysm-related deaths within the 12-month follow-up, and freedom from aneurysm-related mortality was 100% through 1 year.

CONCLUSIONS:

The safety and effectiveness of the GORE EXCLUDER Conformable AAA Endoprosthesis system has been demonstrated with 98.5% freedom from primary effectiveness end point events at 1 year and 100% freedom from primary safety end point events assessed through 30 days.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies Limite: Aged / Aged80 / Humans / Middle aged País/Região como assunto: America do norte Idioma: En Revista: J Vasc Surg Assunto da revista: ANGIOLOGIA Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies Limite: Aged / Aged80 / Humans / Middle aged País/Região como assunto: America do norte Idioma: En Revista: J Vasc Surg Assunto da revista: ANGIOLOGIA Ano de publicação: 2022 Tipo de documento: Article