Development of a radiopaque tiltmeter to improve reproducibility for Fowler's position on chest radiography.
Nakamura, Y; Monzen, H; Tamura, M; Kosaka, H; Kijima, K; Nishimura, Y.
Afiliação
  • Nakamura Y; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka 589-8511, Japan; Department of Radiological Technology, University Hospital, Kyoto Prefectural University of Medicine, Kawaramachi-Hirokoji, Kamigyo-ku, Kyoto 602-8566, Japa
  • Monzen H; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka 589-8511, Japan. Electronic address: hmon@med.kindai.ac.jp.
  • Tamura M; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka 589-8511, Japan. Electronic address: mtam@med.kindai.ac.jp.
  • Kosaka H; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka 589-8511, Japan. Electronic address: h-kosaka@med.kindai.ac.jp.
  • Kijima K; Department of Radiological Technology, University Hospital, Kyoto Prefectural University of Medicine, Kawaramachi-Hirokoji, Kamigyo-ku, Kyoto 602-8566, Japan. Electronic address: kkijima@koto.kpu-m.ac.jp.
  • Nishimura Y; Department of Radiation Oncology, Faculty of Medicine, Kindai University, 377-2 Ohno-Higashi, Osaka-Sayama, Osaka 589-8511, Japan. Electronic address: ynishi@med.kindai.ac.jp.
Radiography (Lond) ; 28(4): 912-918, 2022 11.
Article em En | MEDLINE | ID: mdl-35810623
ABSTRACT

INTRODUCTION:

We have developed a novel radiopaque tiltmeter (ROT) that can indicate patient tilt during a radiography examination and display it on X-ray images. This study evaluated the effect of variation of patient tilt on the reproducibility of Fowler's position for chest radiography and the accuracy of the ROT.

METHODS:

We evaluated the reproducibility of Fowler's position based on changes from the first day in the central venous catheter (CVC) tip position and the cardiothoracic ratio (CTR) with and without a digital tiltmeter to verify its efficacy in patients who underwent mobile chest radiography. The ROT contains radiopaque liquid consisting of white barium sulfate solution and oil and has a scale bar of 15°-75° with increments of 15° to indicate ROT tilt. The ROT tilt was increased from 10° to 80° in increments of 10°. We then evaluated (1) the difference between the ROT tilt and the tilt measured with a digital tiltmeter, and (2) the ROT tilt displayed on the X-ray image.

RESULTS:

With regard to reproducibility in Fowler's position, changes in the CVC tip position were 2.8 ± 3.9 mm and 10.7 ± 10.6 mm with and without the tiltmeter, respectively (p < 0.05) and the respective rates of change in the CTR were 0.7% ± 0.6% and 4.0% ± 2.1% (p < 0.05). Differences between the ROT tilt and the tilt measured by the digital tiltmeter were within ±2.5°. All ROT tilts displayed on the X-ray images were recognized exactly as each tilt.

CONCLUSION:

Our novel ROT had the potential to accurately indicate patient tilt during chest radiography, which could be helpful in terms of reproducibility and precise follow-up. IMPLICATIONS FOR PRACTICE Use of the ROT for determination of patient tilt can improve reproducibility in Fowler's position, allowing more accurate serial X-ray imaging.