q value and parent energy optimization using a low-voltage ionization approach increases resolution in linear ion trap mass spectrometry.
J Mass Spectrom
; 57(8): e4876, 2022 Aug.
Article
em En
| MEDLINE
| ID: mdl-35941810
In this work, the isolation step in the linear ion trap was performed using different "q values" conditions at a low collision-induced dissociation (CID) energy leading to the parent ion resolution improvements, reasonably due to better ion energy distribution. According to the results, we obtained a greater resolution and mass accuracy operating in both traditional electrospray and low voltage ionization near the q value = 0.778 and with a CID energy of 10%. This effect was evaluated with low-molecular-mass compounds (skatole and arginine). The proposed optimization yielded a superior instrument performance without adding technological complexity to mass spectrometry analyses.