Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement.
Cho, Sung-Eun; Han, Jungsun; Park, Ju-Hee; Park, Euna; Kim, Geun Young; Lee, Jun Hyung; Yi, Ahram; Lee, Sang Gon; Lee, Eun Hee; Yun, Yeo-Min.
Afiliação
  • Cho SE; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Han J; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Park JH; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Park E; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Kim GY; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Lee JH; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Yi A; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Lee SG; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Lee EH; Department of Laboratory Medicine, GCLabs, Yongin, Korea.
  • Yun YM; Department of Laboratory Medicine, Konkuk University Medical Center, Konkuk University School of Medicine, Seoul, Korea.
Ann Lab Med ; 43(1): 19-28, 2023 01 01.
Article em En | MEDLINE | ID: mdl-36045053
Background: Mass spectrometry methods exhibit higher accuracy and lower variability than immunoassays at low testosterone concentrations. We developed and validated an ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay for quantifying serum total testosterone. Methods: We used an ExionLC UPLC (Sciex, Framingham, MA, USA) system and a Sciex Triple Quad 6500+ (Sciex) MS/MS system in electrospray ionization and positive ion modes with multiple reaction monitoring transitions to evaluate precision, accuracy, linearity, lower limit of quantitation (LLOQ), carryover, ion suppression, stability, and reference intervals. For method comparison, we measured serum testosterone concentrations using this method in 40 subjects whose testosterone concentrations ranged from 0.14 to 55.48 nmol/L as determined using the Architect i2000 immunoassay (Abbott Diagnostics, Abbott Park, IL, USA) and in an additional 160 sera with testosterone concentrations <1.67 nmol/L. Results: The intra- and inter-run precision CVs were <2.81%, and the accuracy bias values were <3.85%, which were all acceptable. The verified linear interval was 0.03-180.84 nmol/L; the LLOQ was 0.03 nmol/L. No significant carryover and ion suppression were observed. The testosterone in serum was stable at 4°C, at -20°C, and after three freeze-thaw cycles. The reference intervals were successfully verified. The correlation was good at testosterone concentrations of 0.14-55.48 nmol/L; however, the Architect assay showed positive percent bias at concentrations <1.67 nmol/L. Conclusions: The UPLC-MS/MS assay shows acceptable performance, with a lower LLOQ than the immunoassay. This method will enable the quantitation of low testosterone concentrations.