Your browser doesn't support javascript.
loading
Titanium implant surface roughness after different implantoplasty protocols: A laboratory study.
Yildiz, Hulya; Bertl, Kristina; Stavropoulos, Andreas.
Afiliação
  • Yildiz H; Department of Periodontology, Faculty of Dentistry, Istanbul Aydin University, Istanbul, Turkey.
  • Bertl K; Division of Oral Surgery, University Clinic of Dentistry, Medical University of Vienna, Vienna, Austria.
  • Stavropoulos A; Department of Periodontology, Faculty of Odontology, Univesity of Malmö, Malmö, Sweden.
Clin Exp Dent Res ; 8(6): 1315-1321, 2022 12.
Article em En | MEDLINE | ID: mdl-36069295
ABSTRACT

OBJECTIVE:

To compare the surface roughness of sandblasted, large grit, acid-etched (SLA) surfaced titanium discs, after implantoplasty (IP) with different combinations of rotating instruments without or with the subsequent use of a silicone polisher.

METHODS:

Titanium discs (n = 12 per group) with an SLA surface were treated with the following IP protocols (1) Tungsten carbide bur sequence from company 1 (Komet Dental) without or with polishing (P) with a silicone polisher (Brownie®), (2) tungsten carbide bur sequence from company 2 (Hager & Meisinger GmbH) without or with P, and (3) diamond bur sequence (125, 40, 15-µm grit) without or with P. Pristine turned (T) and SLA titanium discs were used as negative and positive controls, respectively. Surface roughness measurements were taken with a contact profilometer rendering Ra and Rz values.

RESULTS:

All IP protocols, even without P, resulted in significantly reduced surface roughness compared to the SLA group. The tungsten carbide bur protocols, even without P, resulted in a surface roughness similar to or significantly lower than that in the T group in terms of Ra and Rz , respectively. IP with the diamond bur sequence resulted in a significantly rougher surface compared to that achieved with the carbide burs. In all IP groups, P with a silicone polisher resulted in a significantly smoother surface.

CONCLUSIONS:

IP with dedicated tungsten carbide burs without or with the subsequent use of a silicone polisher resulted in a surface roughness similar to or significantly lower than that of commercially available turned surfaces. IP with a diamond bur sequence required additional polishing to achieve a comparable surface roughness to that of commercially available turned surfaces.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Dentários / Polimento Dentário Idioma: En Revista: Clin Exp Dent Res Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Turquia País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Dentários / Polimento Dentário Idioma: En Revista: Clin Exp Dent Res Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Turquia País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA