Your browser doesn't support javascript.
loading
Defining the distance between the His bundle and first septal perforator: implications for left bundle branch pacing.
Bocchese, Matthew; Gannon, Michael; Patil, Pravin; Van Decker, William; Whitman, Isaac R; Cronin, Edmond M.
Afiliação
  • Bocchese M; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA. Matthew.bocchese@tuhs.temple.edu.
  • Gannon M; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA.
  • Patil P; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA.
  • Van Decker W; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA.
  • Whitman IR; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA.
  • Cronin EM; Department of Medicine, Section of Cardiology, Lewis Katz School of Medicine, Temple University Hospital, 3401 N. Broad St., Parkinson Pavilion, 9th Floor, PA, 19140, Philadelphia, USA.
J Interv Card Electrophysiol ; 66(5): 1113-1117, 2023 Aug.
Article em En | MEDLINE | ID: mdl-36155878
ABSTRACT

BACKGROUND:

Left bundle branch pacing (LBBP) is a developing method of native conduction pacing, but cases of injury to the septal perforator arteries during implantation have been reported. Knowing the distance between the His bundle and the first septal perforator artery can help operators implant LBBP leads more safely.

METHODS:

Using previously performed coronary CT angiography (CCTA) studies, the distance between the His bundle and the first septal perforator was measured.

RESULTS:

A total of 50 CCTA studies were included. The mean distance from the His bundle to the first septal perforator (His-SP) along the line connecting the His bundle to the RV apex (His-RV apex) was 27.17 ± 7.7 mm with a range of 13.0 to 44.7 mm. The distance was greater than 2.0 cm in 84% of patients. To standardize this distance among patients with varying cardiac structures, the ratio between the His-SP distance and the His-RV Apex distance was also measured. The mean His-SPHis-RV Apex was 0.302 and the median was 0.298. Eighty-six percent of patients had a ratio of greater than 0.20.

CONCLUSION:

Using this information, operators can aim to implant LBBP leads within 2.0 cm of the His bundle or 20% of the distance between the His bundle and the RV apex with minimal risk of causing vascular injury.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bloqueio de Ramo / Fascículo Atrioventricular Limite: Humans Idioma: En Revista: J Interv Card Electrophysiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bloqueio de Ramo / Fascículo Atrioventricular Limite: Humans Idioma: En Revista: J Interv Card Electrophysiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos