Sysmex UN2000 detection of protein/creatinine ratio and of renal tubular epithelial cells can be used for screening lupus nephritis.
Chen, Yabin; Zhao, Yuan; Zhang, Zhishan; Cheng, Xiang; Lin, Jie; Li, Jiaming; Wu, Yibo; Lin, Zhen Zhong; Jing, Jing.
Afiliação
  • Chen Y; Clinical Laboratory, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, China.
  • Zhao Y; Department of Clinical Laboratory Medicine, Xijing Hospital, Fourth Military Medical University (Air Force Medical University), 127 Changle West Road, 710032, Xi'an, Shaanxi, China.
  • Zhang Z; Clinical Laboratory, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, China.
  • Cheng X; Department of Clinical Laboratory Medicine, Xijing Hospital, Fourth Military Medical University (Air Force Medical University), 127 Changle West Road, 710032, Xi'an, Shaanxi, China.
  • Lin J; Department of Clinical Laboratory Medicine, Xijing Hospital, Fourth Military Medical University (Air Force Medical University), 127 Changle West Road, 710032, Xi'an, Shaanxi, China.
  • Li J; Clinical Laboratory, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, China.
  • Wu Y; Clinical Laboratory, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, China.
  • Lin ZZ; Clinical Laboratory, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, China.
  • Jing J; Department of Clinical Laboratory Medicine, Xijing Hospital, Fourth Military Medical University (Air Force Medical University), 127 Changle West Road, 710032, Xi'an, Shaanxi, China. sevencee@sina.com.
BMC Nephrol ; 23(1): 328, 2022 10 05.
Article em En | MEDLINE | ID: mdl-36199035
ABSTRACT

OBJECTIVES:

This study is aimed to evaluate if automated urine sediment analysis UN2000 can be used to screen lupus nephritis.

METHODS:

UN2000 was used to examine 160 urine samples from patients with systemic lupus erythematosus and 124 urine samples from Lupus nephritis. The result of protein/creatinine ratio(P/C) and renal tubular epithelial cells (RTEC) were evaluated. With biochemical analysis and microscopic examination as the gold standard, the Kappa consistency test was used to analyze the accuracy of P/C and RTEC. Analysis was to evaluate the accuracy of P/C single item or RTEC single item and both screening lupusnephritis.

RESULTS:

The consistency of P/C and the gold standard, and that of RTEC and the gold standard are respectively strong and good (0.858 vs. 0.673). The specificity, positive predictive value, and coincidence were the highest when P/C ≥ 2 + was set as the only screening standard for lupus nephritis. When the standard was selected between P/C ≥ 2 + or RTEC > 2.8 cells/µl, the sensitivity and negative predictive value were the highest.

CONCLUSION:

UN 2000 can be used to screen lupus nephritis by detecting P/C and RTEC.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nefrite Lúpica / Lúpus Eritematoso Sistêmico Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Humans Idioma: En Revista: BMC Nephrol Assunto da revista: NEFROLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nefrite Lúpica / Lúpus Eritematoso Sistêmico Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Humans Idioma: En Revista: BMC Nephrol Assunto da revista: NEFROLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China