Independent monitor unit verification for dynamic flattened beam plans on the Halcyon linac.
Kubo, Kazuki; Tamura, Mikoto; Matsumoto, Kenji; Otsuka, Masakazu; Monzen, Hajime.
Afiliação
  • Kubo K; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osaka-sayama, Osaka, Japan.
  • Tamura M; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osaka-sayama, Osaka, Japan.
  • Matsumoto K; Department of Radiology Center, Kindai University Hospital, Osaka-sayama, Osaka, Japan.
  • Otsuka M; Department of Radiology Center, Kindai University Hospital, Osaka-sayama, Osaka, Japan.
  • Monzen H; Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osaka-sayama, Osaka, Japan.
J Appl Clin Med Phys ; 24(1): e13807, 2023 Jan.
Article em En | MEDLINE | ID: mdl-36265085
Independent monitor unit verification (MUV) methods for the dynamic beam-flattening (DBF) technique have not been established. The purpose of this study was to clarify whether MU values for the DBF technique can be calculated using in-air and in-water output ratios (Sc and Scp ). Sc and Scp were measured in the DBF mode, and the phantom scatter factor (Sp ) was calculated. The difference between calculated and planned MUs with square and rectangle fields and clinical plans for different treatment sites was also evaluated. Sc values for the 4 × 4 to 24 × 24 cm2 fields of the distal multi-leaf collimator (MLC) layer at 2-cm intervals were 0.887, 0.815, 0.715, 0.716, 0.611, 0.612, 0.511, 0.373, 0.374, 0.375, and 0.374, respectively. No collimator exchange effect was observed. Sc also depends slightly on the field size of the distal MLC layer. If the distal-MLC-layered field size was less than 20% of the corresponding MLC sequence size in the proximal MLC layer, Sc was affected by >1%, which was compensated using a correction factor (CF). Sp increased as the field sizes of the MLC sequence and distal MLC leaves increased. MUs calculated using measured Sc , Sp , and CF for square and rectangle fields agreed with planned MUs within ±1.2%. A larger difference (-1.5%) between calculated and planned MUs was observed for clinical plans, whereas differences in MUs were within 2 MU for most fields (56 out of 64 fields). MU calculation for the DBF technique can be performed with Sc , Sp , and CF for independent MUV.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Planejamento da Radioterapia Assistida por Computador Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Planejamento da Radioterapia Assistida por Computador Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão País de publicação: Estados Unidos