A Reamer by Another Number: What Do They Actually Measure?
Alexander, Jacob S; Morris, Michael J; Lombardi, Adolph Jr V; Berend, Keith R; Crawford, David A.
Afiliação
  • Alexander JS; JIS Orthopedics, New Albany, Ohio.
  • Morris MJ; JIS Orthopedics, New Albany, Ohio.
  • Lombardi AJV; JIS Orthopedics, New Albany, Ohio.
  • Berend KR; JIS Orthopedics, New Albany, Ohio.
  • Crawford DA; JIS Orthopedics, New Albany, Ohio.
Surg Technol Int ; 412022 Sep 22.
Article em En | MEDLINE | ID: mdl-36326046
Acetabular reamers are a mainstay in the preparation of the acetabulum in total hip arthroplasty. Many surgeons, however, have noticed a discrepancy in the preparation of the socket and the reported size of reamers during this crucial phase of the procedure. A cross-sectional study performed by measuring a variety of sets of acetabular reamers was performed. The results showed that 17 of 21 sizes (81%) measured at least 2mm less than their reported size. Given that these differences beget unintended consequences, the industry may consider replacing or sharpening reamers already in use or even switching to a single-use reamer model.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies Idioma: En Revista: Surg Technol Int Ano de publicação: 2022 Tipo de documento: Article País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies Idioma: En Revista: Surg Technol Int Ano de publicação: 2022 Tipo de documento: Article País de publicação: Estados Unidos