Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility.
Decker, Josua A; O'Doherty, Jim; Schoepf, U Joseph; Todoran, Thomas M; Aquino, Gilberto J; Brandt, Verena; Baruah, Dhiraj; Fink, Nicola; Zsarnoczay, Emese; Flohr, Thomas; Schmidt, Bernhard; Allmendinger, Thomas; Risch, Franka; Varga-Szemes, Akos; Emrich, Tilman.
Afiliação
  • Decker JA; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • O'Doherty J; Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Augsburg, Germany.
  • Schoepf UJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • Todoran TM; Siemens Medical Solutions, Malvern, PA, USA.
  • Aquino GJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA. schoepf@musc.edu.
  • Brandt V; Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, SC, USA.
  • Baruah D; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • Fink N; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • Zsarnoczay E; Department of Cardiology and Angiology, Robert-Bosch Hospital, Stuttgart, Germany.
  • Flohr T; Division of Thoracic Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Schmidt B; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • Allmendinger T; Department of Radiology, University Hospital, LMU Munich, Munich, Germany.
  • Risch F; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Drive, Charleston, SC, USA.
  • Varga-Szemes A; Medical Imaging Centre, Semmelweis University, Budapest, Hungary.
  • Emrich T; Siemens Healthcare GmbH, Forchheim, Germany.
Eur Radiol ; 33(4): 2469-2477, 2023 Apr.
Article em En | MEDLINE | ID: mdl-36462045
OBJECTIVES: To assess the impact of scan modes and reconstruction kernels using a novel dual-source photon-counting detector CT (PCD-CT) on lumen visibility and sharpness of different stent sizes. METHODS: A phantom containing six balloon-expandable stents (2.5 to 9 mm diameter) in silicone tubing was scanned on a PCD-CT with standard (0.6 mm and 0.4 mm thicknesses) and ultra-high-resolution (0.2 mm thickness) modes. With the use of increasing contrast medium concentrations, densities of 0, 200, 400, and 600 HU were achieved. Standard-resolution scans were reconstructed using increasing sharpness kernels, using both polyenergetic quantitative soft tissue "conventional" ((Qr40c(0.6 mm), Qr40c(0.4 mm), Qr72c(0.2 mm)) and vascular (Bv) virtual monoenergetic reconstructions (Bv44m(0.4 mm), Bv60m(0.4 mm)) at 70 keV. In-stent lumen visibility, sharpness (max. ΔHU of the stent measured in profile plots), and in-stent noise (standard deviation of HU) were measured. RESULTS: In-stent lumen visibility was highest for Qr72c(0.2 mm) (86.5 ± 2.8% to 88.3 ± 2.6%) and in Bv60m(0.4 mm) reconstructions (77.3 ± 2.9 to 82.7 ± 2.5%). Lumen visibility was lowest in the smallest stent (2.5 mm) ranging from 54.1% in Qr40c(0.6 mm) to 74.1% in Qr72c(0.2 mm) and highest in the largest stent (9 mm) ranging from 93.8% in Qr40c(0.6 mm) to 99.1% in the Qr72c(0.2 mm) series. Lumen visibility decreased by 2.1% for every 200-HU increase in lumen attenuation. Max. ΔHU between stents and stent lumen was highest in Qr72c(0.2 mm) (ΔHU 892 ± 504 to 1526 ± 517) and Bv60m(0.4 mm) series (ΔHU 480 ± 357 to 1030 ± 344). Improvement of lumen visibility and sharpness in UHR and Bv60m(0.4 mm) series was strongest in smaller stent sizes. CONCLUSION: UHR acquisition mode and sharp reconstruction kernels on a novel PCD-CT system significantly improve in-stent lumen visibility and sharpness-especially for smaller stent sizes. KEY POINTS: • In-stent lumen visibility and sharpness of stents significantly improve using sharp reconstruction kernels (Bv60) and ultra-high-resolution mode in photon-counting detector computed tomography. • The observed improvement of stent-lumen visibility was highest in smaller stent sizes.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Stents Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Stents Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Alemanha