Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility.
Eur Radiol
; 33(4): 2469-2477, 2023 Apr.
Article
em En
| MEDLINE
| ID: mdl-36462045
OBJECTIVES: To assess the impact of scan modes and reconstruction kernels using a novel dual-source photon-counting detector CT (PCD-CT) on lumen visibility and sharpness of different stent sizes. METHODS: A phantom containing six balloon-expandable stents (2.5 to 9 mm diameter) in silicone tubing was scanned on a PCD-CT with standard (0.6 mm and 0.4 mm thicknesses) and ultra-high-resolution (0.2 mm thickness) modes. With the use of increasing contrast medium concentrations, densities of 0, 200, 400, and 600 HU were achieved. Standard-resolution scans were reconstructed using increasing sharpness kernels, using both polyenergetic quantitative soft tissue "conventional" ((Qr40c(0.6 mm), Qr40c(0.4 mm), Qr72c(0.2 mm)) and vascular (Bv) virtual monoenergetic reconstructions (Bv44m(0.4 mm), Bv60m(0.4 mm)) at 70 keV. In-stent lumen visibility, sharpness (max. ΔHU of the stent measured in profile plots), and in-stent noise (standard deviation of HU) were measured. RESULTS: In-stent lumen visibility was highest for Qr72c(0.2 mm) (86.5 ± 2.8% to 88.3 ± 2.6%) and in Bv60m(0.4 mm) reconstructions (77.3 ± 2.9 to 82.7 ± 2.5%). Lumen visibility was lowest in the smallest stent (2.5 mm) ranging from 54.1% in Qr40c(0.6 mm) to 74.1% in Qr72c(0.2 mm) and highest in the largest stent (9 mm) ranging from 93.8% in Qr40c(0.6 mm) to 99.1% in the Qr72c(0.2 mm) series. Lumen visibility decreased by 2.1% for every 200-HU increase in lumen attenuation. Max. ΔHU between stents and stent lumen was highest in Qr72c(0.2 mm) (ΔHU 892 ± 504 to 1526 ± 517) and Bv60m(0.4 mm) series (ΔHU 480 ± 357 to 1030 ± 344). Improvement of lumen visibility and sharpness in UHR and Bv60m(0.4 mm) series was strongest in smaller stent sizes. CONCLUSION: UHR acquisition mode and sharp reconstruction kernels on a novel PCD-CT system significantly improve in-stent lumen visibility and sharpness-especially for smaller stent sizes. KEY POINTS: ⢠In-stent lumen visibility and sharpness of stents significantly improve using sharp reconstruction kernels (Bv60) and ultra-high-resolution mode in photon-counting detector computed tomography. ⢠The observed improvement of stent-lumen visibility was highest in smaller stent sizes.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Tomografia Computadorizada por Raios X
/
Stents
Limite:
Humans
Idioma:
En
Revista:
Eur Radiol
Assunto da revista:
RADIOLOGIA
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
Estados Unidos
País de publicação:
Alemanha