Your browser doesn't support javascript.
loading
Clinical Outcome of Transcatheter Aortic Valve Replacement With TriGUARD 3™ Cerebral Embolic Protection Device.
Daal, Sayonara M; Jimenez-Rodriguez, Gian M J; Voskuil, Michiel; Kraaijeveld, Adriaan O; Dessing, Thomas C; Ramjankhan, Faiz Z; Mokhles, Mostafa M; Stella, Pieter R.
Afiliação
  • Daal SM; Department of Cardiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands; Department of Cardiac Surgery, Haga Teaching Hospitals, The Hague, the Netherlands. Electronic address: sayonara.daal@hotmail.com.
  • Jimenez-Rodriguez GMJ; Department of Cardiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: gianma21@hotmail.com.
  • Voskuil M; Department of Cardiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: mvoskuil@umcutrecht.nl.
  • Kraaijeveld AO; Department of Cardiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: a.o.kraaijeveld-3@umcutrecht.nl.
  • Dessing TC; Department of Cardiothoracic Surgery, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: t.c.dessing@umcutrecht.nl.
  • Ramjankhan FZ; Department of Cardiothoracic Surgery, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: f.ramjankhan@umcutrecht.nl.
  • Mokhles MM; Department of Cardiothoracic Surgery, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: m.m.mokhles@umcutrecht.nl.
  • Stella PR; Department of Cardiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, the Netherlands. Electronic address: pieter.stella@gmail.com.
Cardiovasc Revasc Med ; 50: 8-12, 2023 05.
Article em En | MEDLINE | ID: mdl-36681530
ABSTRACT

OBJECTIVE:

Periprocedural stroke during transcatheter aortic valve replacement (TAVR) is a highly feared adverse event. The TriGUARD 3 cerebral embolic protection device (CEPD) may have the potential benefit of reduction of embolic events, but it still remains unclear whether it reduces the incidence of periprocedural stroke or transient ischemic attack (TIA). We aimed to investigate whether the latest TriGUARD 3 CEPD reduces the incidence of clinically overt stroke within 72 h or at discharge after TAVR.

METHODS:

In this prospective single-center study 117 patients (mean age 80.3 years, 53.8 % male) were included from July 2020 to December 2021.

RESULTS:

The primary efficacy endpoint of this study, periprocedural clinically overt stroke or TIA, within 72 h or at discharge after TAVR with the TriGUARD 3 CEPD occurred in 1/117 pts (0.8 %). Secondary endpoints (device related issues such as life-threatening or disabling bleeding, acute kidney injury, major vascular complications) were reported in 4/117 pts (3.4 %).

CONCLUSIONS:

This study suggests that the use of the latest TriGUARD 3™ CEPD in transfemoral TAVR seems to be associated with a low rate of clinically overt stroke and a low rate of device related adverse events, reflecting "real world" TAVR practice. However these results should be hypothesis generating and confirmed in a large RCT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Ataque Isquêmico Transitório / Acidente Vascular Cerebral / Embolia Intracraniana / Dispositivos de Proteção Embólica / Substituição da Valva Aórtica Transcateter Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged80 / Female / Humans / Male Idioma: En Revista: Cardiovasc Revasc Med Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Ataque Isquêmico Transitório / Acidente Vascular Cerebral / Embolia Intracraniana / Dispositivos de Proteção Embólica / Substituição da Valva Aórtica Transcateter Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged80 / Female / Humans / Male Idioma: En Revista: Cardiovasc Revasc Med Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2023 Tipo de documento: Article